Issue title: Frontiers in Biomedical Engineering and Biotechnology – Proceedings of the 2nd International Conference on Biomedical Engineering and Biotechnology, 11–13 October 2013, Wuhan, China
Article type: Research Article
Authors: Liu, Guangmao | Zhang, Yan | Chen, Haibo | Sun, Hansong | Zhou, Jianye; | Hu, Shengshou
Affiliations: State Key Laboratory of Cardiovascular Disease, Key Laboratory of Cardiovascular Regenerative Medicine, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, People's Republic of China
Note: [] Address for correspondence: Jianye Zhou. Email: [email protected]
Abstract: The properties of blood flow in the outflow cannula of an axial blood pump play a critical role in potential thrombus formation and vascular injury. In this study, an in vitro flow visualization technique using particle image velocimetry (PIV) was applied to investigate the flow characteristics in the outflow cannula of a FW-2 model axial pump. The two-dimensional (2-D) flow field in the axial central section and the three-dimensional (3-D) flow field in the whole outflow cannula were examined with the PIV system. Tests were carried out with a blood-mimic working fluid in the axial pump at a rotational speed of 8500±20 rpm with a flow rate of 5 L/min. The velocity distribution in the outflow cannula was analyzed to evaluate the flow characteristics. There was no backflow or stagnant flow in the tested area, while the flow velocity rapidly increased outside the boundary layer. A spiral flow was observed near the boundary layer, but this was worn off within the tested area. Based on the results, hemolysis and thrombus formation in the cannula, and injury to aortic endothelium are unlikely to occur due to spiral flow.
Keywords: particle image velocimetry, axial blood pump, outflow cannula, thrombus formation
DOI: 10.3233/BME-130791
Journal: Bio-Medical Materials and Engineering, vol. 24, no. 1, pp. 117-122, 2014
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]