Article type: Research Article
Authors: Takahashi, Toshiyuki; | Tominaga, Teiji | Yoshimoto, Takashi | Koshu, Keiji | Toshimitsu Yokobori, Jr, A. | Aizawa, Yoichi
Affiliations: Department of Neurosurgery, Tohoku University School of Medicine, 1-1 Seiryo-machi, Aoba-ku, Sendai 980-77, Japan | Department of Neurosurgery, Kohnan Hospital, 4-20-1 Nagamachi-Minami, Taihaku-ku, Sendai 982, Japan | Department of Mechatronics and Precision Engineering, Tohoku University, Aoba Aramaki, Aoba-ku, Sendai 980-77, Japan
Note: [] Address for reprint requests to: T. Takahashi, Department of Neurosurgery, Tohoku University School of Medicine, 1-1 Seiryo-machi, Aoba-ku, Sendai 980-77, Japan. Tel.: +81 22 717 7230; Fax: +81 22 717 7233.
Abstract: The biomechanical properties of a porous hydroxyapatite (HA) intervertebral graft with or without anterior cervical plating were evaluated in cadaveric porcine cervical spine model using C3–4 discectomy and dissection of the posterior longitudinal ligament to cause instability. The experimental groups were intact (n = 11), instability (n = 11), autogenous bone implant (n = 6), HA graft implant (n = 5), autogenous bone with plating (n = 6), and HA graft with plating (n = 5). Porous HA with 40% porosity and scapular bone were used as grafts. The displacement rates of the cervical spine by compressive forces in the flexural, extensional, and lateral bending directions were evaluated using video-recording followed by computer-assisted analysis. The stiffness to compressive load was calculated from the load-displacement rate curve. The linear and non-linear coefficients of the Fung's equation were obtained based on the plot of Young's modulus against load. There were no statistical differences in the stiffness between the HA and autogenous bone graft in all directions. The two plating groups showed significantly increased stiffness in all directions. The non-linear coefficient value in Fung's equation was far larger in both HA and HA graft with plating groups than in the other groups in flexural compression. Porous HA graft has a compressive strength similar to autogenous graft in vitro, and anterior plating provides additional stiffness to the cervical spine. The larger non-linear coefficient value of the HA groups may represent the characteristic biomechanical brittleness of HA graft, but this is manifest only in flexural compression.
Keywords: Cervical spine, anterior fixation, instrumentation, hydroxyapatite, biomechanics
DOI: 10.3233/BME-1997-7204
Journal: Bio-Medical Materials and Engineering, vol. 7, no. 2, pp. 121-127, 1997
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]