Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Niina, Onnela; | Hirotaka, Takeshita | Yoshiuki, Kaiho | Toshiya, Kojima | Risato, Kobayashi | Tetsu, Tanaka | Jari, Hyttinen
Affiliations: Department of Biomedical Engineering, Tampere University of Technology, Tampere, Finland | Department of Bioengineering and Robotics, Tohoku University, Sendai, Japan | Technical Division, School of Engineering, Tohoku University, Sendai, Japan
Note: [] Address for correspondence: Onnela Niina, Department of Biomedical Engineering, Tampere University of Technology, Korkeakoulunkatu 3, 33720 Tampere, Finland. Tel.: +358 40 849 0024; Fax: +358 3 3115 2162; E-mail: [email protected].
Abstract: Retinal implants may provide vision for people suffering from photoreceptor degeneration caused by different eye diseases. Electrode size in retinal implant should be decreased in order to increase the resolution provided by the implant. We defined electric properties of five different electrode materials (Au, Ir-b, Ti, TiN, Pt-b) widely used in retinal prostheses. The comparison of different electrode materials requires that the electrical properties of different materials are defined using exactly the same measurement conditions and devices. Existing studies about electrode material properties are often made using slightly different measurement parameters or electrode processing conditions making the comparison between different materials difficult. Here, the electrochemical characterization included cyclic voltammetry and electrochemical impedance spectroscopy. Ir-b and Pt-b had greater charge injection capacity than other materials. The fabricated material samples showed that in this experiment the electrode diameter larger than 200 μm should be used to suppress irreversible reaction of stimulus electrodes with the needed stimulus currents. Thus, either we have to find novel electrode materials or surface treatment methods to decrease the electrode area providing increased electrode and pixel number of the prosthesis or we have to show that stimulus currents smaller than 40 μA are enough to induce phosphenes.
Keywords: Retina prosthesis, cyclic voltammetry, electrical impedance spectroscopy, electrical properties of electrodes
DOI: 10.3233/BME-2011-0658
Journal: Bio-Medical Materials and Engineering, vol. 21, no. 2, pp. 83-97, 2011
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]