Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Grosland, Nicole | Kim, Jeong K. | Park, Joon B.
Affiliations: Department of Biomedical Engineering, College of Engineering, The University of Iowa, Iowa City, IA 52242, USA
Abstract: The longevity of cemented total joint replacements, namely total hip arthroplasty (THA), depends on the integrity of the cement per se and interfaces of the metallic stem-cement and bone-cement, and stress/strain transmission and its distribution. A simple coil made of stainless steel wire was placed around a tapered stem in a simulated cement mantle to counteract the radial- and hoop-stresses created by the stem loading. Two-dimensional axisymmetric finite element analysis was performed to elucidate the level of the stress and its distribution with and without the wire coil. The results suggest that the peak stresses distributed throughout the cement may be reduced by more than one half, via the incorporation of a wire coil, confirming the mechanical test results, which showed an average fracture load of 3.70 ± 1.13 kN for pure bone cement control and 9.02 ± 1.54 kN for the wire coil reinforced specimens. These results indicate that the reinforcement of the cement mantle with a wire coil in the distal end of the hip stem prosthesis could reduce the hoop-stresses, which may results in reduced stresses in the bone and cement interface. Consequently, loosening of the interfaces of the bone-cement and cement-stem could be reduced due to the enhanced fatigue life of the bone cement mantle with favorable stress distribution at the distal tip of the stem.
Keywords: finite element model, bone cement, fixation, wire coil reinforcement, mechanical properties
DOI: 10.3233/BME-1995-5104
Journal: Bio-Medical Materials and Engineering, vol. 5, no. 1, pp. 29-36, 1995
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]