Article type: Research Article
Authors: Murdoch, A.H.; | Mathias, K.J. | Shepherd, D.E.T.
Affiliations: Department of Bio‐Medical Physics and Bio‐Engineering, University of Aberdeen, Scotland, UK | School of Engineering, Mechanical Engineering, University of Birmingham, Edbaston, Birmingham, UK
Note: [] Corresponding author: Dr A.H. Murdoch, Department of Bio‐Medical Physics and Bio‐Engineering, University of Aberdeen, Foresterhill, Aberdeen, AB25 2ZD, UK. Tel.: 01224 552746; Fax: 01224 685645; E‐mail: [email protected].
Abstract: When testing medical implants it is very important to be able to test the implant using a suitable material. In the case of orthopaedic implants the optimum material is bone. Beech wood is considered a suitable substitute for bone as it has a similar Young's modulus in tension. Although it is widely used, no actual comparison of the two materials has been undertaken. The aim of this study was to compare the material properties of beech wood and cortical bone using conventional compression tests. Cortical bone samples 4 mm in diameter and 20 mm in length, were prepared from the tibia of an amputated leg. Beech wood samples were prepared to the same specifications. In compression, the Young's modulus for cortical bone was found to be 27±9.9 GPa (mean ± standard deviation) and for beech wood 2.6±1.7 GPa. The failure load for cortical bone was 911±207 N and 732±62 N for beech wood. Although beech wood has been used as a substitute for bone in some studies, this study has shown that there are significant differences in the properties of the two materials when they are subjected to compression.
Keywords: Beech wood, cortical bone, compression, Young's modulus
Journal: Bio-Medical Materials and Engineering, vol. 14, no. 1, pp. 1-4, 2004
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]