Article type: Research Article
Authors: Kamangar, Sarfaraza; * | Badruddin, Irfan Anjuma | Ameer Ahamad, N.b | Soudagar, Manzoor Elahi M.c | Govindaraju, Kalimuthud | Nik-Ghazali, N.a | Salman Ahmed, N.J.e | Yunus Khan, T.M.a
Affiliations: [a] Department of Mechanical Engineering, University of Malaya, Kuala Lumpur 50603, Malaysia | [b] Mathematics Department, Faculty of Science, University of Tabuk, Saudi Arabia | [c] Department of Mechanical Engineering, Jain College of Engineering, Belagavi, 590014, India | [d] Department of Mechanical and Industrial Engineering, Mekelle University, Mekelle, Ethiopia | [e] Department of Mechanical and Industrial Engineering, Sultan Qaboos University, Al Khoud, Muscat, 123, Oman
Correspondence: [*] Corresponding author. Tel.: 603-79674463; Fax: 603-79675317; E-mail: [email protected].
Abstract: The current study investigates the effect of multi stenosis on the hemodynamic parameters such as wall pressure, velocity and wall shear stress in the realistic left coronary artery. Patients CT scan image data of normal and diseased left coronary artery was chosen for the reconstruction of 3D coronary artery models. The diseased 3D model of left coronary artery shows a narrowing of more than 70% and 80% of area stenosis (AS) at the left main stem (LMS) and left circumflex (LCX) respectively. The results show that the decrease in pressure was found downstream to the stenosis as compared to the coronary artery without stenosis. The maximum pressure drop was noted across the 80% AS at the left circumflex branch. The recirculation zone was also observed immediate to the stenosis and highest wall shear stress was found across the 80% area stenosis. Our analysis provides an insight into the distribution of wall shear stress and pressure drop, thus improving our understanding on the hemodynamics in realistic coronary artery.
Keywords: Coronary artery, non-Newtonian flow, stenosis, CFD, computer tomography
DOI: 10.3233/BME-171672
Journal: Bio-Medical Materials and Engineering, vol. 28, no. 3, pp. 257-266, 2017
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]