Article type: Research Article
Authors: Endo, M.M. | Barbour, P.S.M. | Barton, D.C. | Wroblewski, B.M. | Fisher, J.; | Tipper, J.L. | Ingham, E. | Stone, M.H.
Affiliations: Department of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, UK | Department of Microbiology, University of Leeds, Leeds LS2 9JT, UK | Department of Orthopaedic Surgery, Leeds General Infirmary, Leeds, UK
Note: [] Address correspondence to: Prof. J. Fisher, School of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, UK.
Abstract: The wear debris generated from UHMWPE (ultra high molecular weight polyethylene) has been recognised as one of the major causes of failure in THR (total hip replacement). GUR 1120 (compression moulded) and GUR 4150HP (ram extruded) which are currently the most frequently used materials in THR were studied in pin‐on‐plate wear test. The wear particles generated from this test were observed by scanning electron micrograph and analysed by image analysis. The results from this study showed that GUR 4150HP had superior wear resistance than GUR 1120 under relatively high wear factor conditions. These results also highlighted the importance of multidirectional motion and its effect on the wear rates of UHMWPE. The multidirectional motion tended to show a higher wear factor than previous studies using unidirectional motion conducted under otherwise similar conditions. The wear debris analysis conducted with the wear particles collected from unidirectional (relatively rough) pin‐on‐plate wear tests (GUR 1120 and GUR 4150HP) showed that the greatest number of particles had a size range of 0.1–0.5 \mum followed by 0.5–1.0 \mum, 1.0–5.0 \mum and 5.0–10.0 \mum, in both GUR 1120 and GUR 4150HP. However, comparing the masses of the wear particles, the bigger size range of greater than 10 \mum, had the highest percent mass followed by 1.0–5.0 \mum, 0.5–1.0 \mum, 0.1–0.5 \mum and 5.0–10.0 \mum.
Journal: Bio-Medical Materials and Engineering, vol. 9, no. 2, pp. 113-124, 1999
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]