Article type: Research Article
Authors: Park, Chan-Hee | Tijing, Leonard D. | Yun, Yeoheung | Kim, Cheol Sang; ;
Affiliations: Department of Bionanosystem Engineering, Chonbuk National University, Jeonju, Jeonbuk, Korea | Division of Mechanical Design Engineering, Chonbuk National University, Jeonju, Jeonbuk, Korea | Department of Bioengineering, North Carolina A&T State University, Greensboro, NC, USA
Note: [] Address for correspondence: C.S. Kim, Division of Mechanical Design Engineering, Chonbuk National University, Jeonju, Jeonbuk 561-756, Korea. Tel.: +82 63 270 4284; Fax: +82 63 270 2460; E-mail: [email protected].
Abstract: This article describes a preliminary investigation and prototype fabrication of a novel potential sensing method to continuously monitor vascular stent fractures. A potential measurement system consisting of Wheatstone bridge circuit and signal conditioning circuit was designed for the cardiovascular stent durability and fatigue test. Each end of a bare and polyurethane-covered Nitinol vascular stent was electrically connected to the potential measurement system and then immersed either in simulated body fluid (SBF) media or distilled water at 36.4±1°C. When the stent experienced fracture (i.e., a cut), its electrical potential decreased with an increase in electrical resistance. This method successfully measured fractures in the stent regardless of location. Furthermore, the number of cycles at the onset of stent fracture was accurately detected and continuously monitored using this technique. Thus, the present fracture detection method, which to our knowledge is the first ever report to use electrical potential measurement for stent durability test, gives a fast, real-time, accurate and efficient detection of fractures in stent during in vitro fatigue and durability test.
Keywords: , Electrical potential, stent fracture, fatigue and durability test, vascular stent, Nitinol, Wheatstone bridge circuit
DOI: 10.3233/BME-2011-0670
Journal: Bio-Medical Materials and Engineering, vol. 21, no. 4, pp. 213-222, 2011
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]