Article type: Research Article
Authors: Chen, Lih-Huei; | Lai, Po-Liang; | Tai, Ching-Lung | Niu, Chi-Chien; | Fu, Tsai-Sheng; | Chen, Wen-Jer; ;
Affiliations: Department of Orthopaedic Surgery, Chang Gung Memorial Hospital, Kweishan, Taoyuan, Taiwan | Chang Gung University, Kweishan, Taoyuan, Taiwan | Biomechanical Laboratory, Chang Gung Memorial Hospital, Kweishan, Taoyuan, Taiwan
Note: [] Corresponding author: Wen-Jer Chen, MD, Department of Orthopaedic Surgery, Chang Gung Memorial Hospital, 5, Fu-Hsing St. 333, Kweishan, Taoyuan, Taiwan. Tel.: +866 3 3281200, ext. 2163; Fax: +886 3 3278113; E-mail: [email protected].
Abstract: Eight fresh porcine lumbar spines received a posterior instrumentation at L4–L5 using pedicle screw–rod system. Each specimen was tested utilizing laminectomies of varying extent. Group A (Integrity) preserved the spinous process and interspinous ligament; Group B (Partial laminectomy) removed the inferior portion of L4 spinous process and preserved the interspinous ligament of L3–L4; Group C (Complete laminectomy) removed the entire L4 spinous process. Hydraulic testing machine was used to generate an increasing moment up to 8400 N mm in flexion and extension. The intervertebral displacement on the superior adjacent disc between L3–L4 was measured using an extensometer. Under extension, no significant difference in the intervertebral displacement was observed among three different models of laminectomy. However, under flexion, the intervertebral displacement on adjacent disc with complete laminectomy was statistically larger than those of integrity and partial laminectomies (P=0.000976 and P=0.0363, respectively). No difference was found between integrity and partial laminectomy groups (P>0.05). This study implies that an instrumented spine with integrity of posterior complex is less likely to develop adjacent instability than a spine with destruction of the anchoring point for supraspinous ligament.
Keywords: Spinal fusion, adjacent segment, interspinous ligament, pedicle screw fixation, porcine model
Journal: Bio-Medical Materials and Engineering, vol. 16, no. 4, pp. 261-267, 2006
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]