Issue title: A Selection of Papers from the International Symposium on Advanced Bio-Materials and Engineering '93 (ISABE '93)
Subtitle: Part II: The Implant/Bone Interface
Article type: Research Article
Authors: Albert, K. | Schledjewski, R. | Harbaugh, M. | Bleser, S.; | Jamison, R. | Friedrich, K.
Affiliations: Smith & Nephew Richards Inc., 1450 Brooks Road, Memphis, Tennessee, USA | Institute for Composite Materials, University of Kaiserslautem, Kaiserslautem, Germany
Abstract: Carbon fiber/PEEK polymer (C/PEEK) composite materials are being developed for use as orthopaedic implant materials, Wear is an issue of increasing importance in orthopaedic implants; particulate debris generated by the wearing of biomaterials may be a causal factor leading to osteolysis and implant loosening, Therefore, numerical and ellperimental studies were completed to characterize the wear of C/PEEK composite materials in comparison to current orthopaedic implant materials. Finite element analyses (FEA) of a composite material hip stem implanted in a femur and loaded at 890 N determined that peak contact stresses will occur at the proximal-medial and distal regions of the implant. These contact stresses were found to be below 1.0 MPa over most of the implant surface; however the peak stress in the proximal-medial region was 1.8 MPa and higher still at the distal portion of the stem, In vivo forces result in contact stress values up to 9.0 MPa, The composite implant exhibited 10–40% lower contact stresses in the distal region compared to a titanium-alloy implant of identical design. Composite material wear samples were slid against porous hydroxylapatite (HA) to simulate the stem/bone interface. An identical series of experiments was run for comparison to a current orthopaedic implant material—Ti6Al4V titanium alloy. Two domains of motion were studied; a composite ring-on-HA disc large amplitude sliding wear test; and a composite pin-on-HA disc small amplitude fretting regimen. Nominal contact pressures during testing were 1.4 MPa and 7.6 MPa for sliding and fretting tests, respectively. Fretting and sliding abrasive wear tests resulted in the composite material exhibiting a lower wear rate than the titanium-alloy. The magnitude of the difference was greatly dependent on the contact pressures, sliding amplitudes, and counterface material properties.
Keywords: composite material hip, implant, wear, C/PEEK
DOI: 10.3233/BME-1994-4308
Journal: Bio-Medical Materials and Engineering, vol. 4, no. 3, pp. 199-211, 1994
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]