Article type: Research Article
Authors: Zhang, Haifena; | Lai, Lanshuhuia | Zhao, Shiyua | Liu, Qiana | Fang, Ninaa | Zhao, Lia | Yang, Yana
Affiliations: [a] Department of Nursing, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, Shanghai, China
Correspondence: [*] Corresponding author: Haifen Zhang, Department of Nursing, Renji Hospital, School of Medicine, Shanghai Jiao Tong University, No. 160 Pujian Road, Shanghai 200127, China. Tel.: +86 21 58752345; Fax: +86 21 58394262; E-mail: [email protected]
Abstract: BACKGROUND:Portable blood glucose meters are the main method for detecting the blood glucose status of clinical patients. OBJECTIVE:To investigate the accuracy of detecting blood glucose in haemodialysis patients by sampling two blood glucose meters through the haemodialysis line. METHODS:Convenient sampling was used to select 80 patients with maintenance haemodialysis. The patients were sampled through the arterial end of the haemodialysis line within three minutes of being put on the machine. One specimen was tested by glycemeter1, which can identify the type of blood in the arteries and veins, and glycemeter2, which can only detect blood glucose in the capillaries for bedside blood glucose testing. The other specimen was sent to the laboratory biochemical analyser for blood glucose testing. RESULTS:When the blood glucose value of the first blood glucose meter (No. 1) was compared with the laboratory biochemical analyser, the correlation coefficient was r = 0.805 (p < 0.05), the out of value of the first blood glucose meter accounted for 4.4%, and the consistency reached 95% (p < 0.05). When the blood glucose value of the second blood glucose meter (No. 2) was compared with the laboratory biochemical analyser, the correlation coefficient was r = 0.800 (p < 0.05), the out of value of the second blood glucose meter accounted for 4.4%, and the consistency reached 95% (p < 0.05). CONCLUSIONS:For patients with maintenance haemodialysis, the blood glucose values detected by the two bedside blood glucose meters using arteriovenous mixed blood in the pipeline do not affect the accuracy and can respond more realistically.
Keywords: Blood glucose meter, arteriovenous mixed blood, whole blood glucose, plasma glucose, hemodialysis
DOI: 10.3233/BME-211268
Journal: Bio-Medical Materials and Engineering, vol. 33, no. 2, pp. 113-121, 2022
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]