Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Matsumoto, Takeo | Naiki, Takeru | Hayashi, Kozaburo;
Affiliations: Department of Scientific Instrumentation and Control, Research Institute for Electronic Science, Hokkaido University, Sapporo 060, Japan | Department of Mechanical Engineering, Faculty of Engineering Science, Osaka University, Toyonaka, Osaka 560, Japan
Abstract: Flow characteristics near the end-to-end anastomosis of vascular graft were studied in model tubes by flow visualization techniques. Artery and vascular graft were modelled by an elastic tube fabricated from an elastomeric polymer and a rigid plastic tube, respectively. Anastomotic models were made by connecting these two tubes, which had compliance mismatch at their anastomoses. These model tubes were installed into a mock circulatory loop and flow was visualized using hydrogen bubbles and aluminum powder as the tracer. Flow disturbances including flow separation and eddies were observed near the modelled distal anastomosis (graft-to-artery anastomosis). Peak values of the wall shear rate were high in the proximal anastomotic area (artery-to-graft anastomosis) and low in the distal region. These phenomena were enhanced in the models with increased compliance mismatch. The local abnormal flow observed in the anastomotic zone might cause thrombus formation and subintimal hyperplasia. To improve the patency in small-calibered arterial grafts, it is important to match their compliance to that of natural arteries.
Keywords: arterial prosthesis, compliance mismatch, end-to-end anastomosis, flow visualization, hydrogen bubble method
DOI: 10.3233/BME-1992-2401
Journal: Bio-Medical Materials and Engineering, vol. 2, no. 4, pp. 171-183, 1992
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]