Article type: Research Article
Authors: Zhang, Zhenjuna; b | Fogel, Guy R.c | Liao, Zhenhuab | Sun, Yitaod | Sun, Xuejune | Liu, Weiqianga; b;
Affiliations: [a] Department of Mechanical Engineering, Tsinghua University, Beijing, China | [b] Biomechanics and Biotechnology Lab, Research Institute of Tsinghua University in Shenzhen, Shenzhen, China | [c] Spine Pain Begone Clinic, San Antonio, TX, USA | [d] Haicheng City Central Hospital, Haicheng, China | [e] Beijing Yingnuanlihe Technologies Co., Ltd, Beijing, China
Correspondence: [*] Corresponding author: Weiqiang Liu, Tsinghua University, Haidian District, Beijing, China. Tel.: +86 0755 26551376; Fax: +86 0755 26551380; E-mail: [email protected]
Abstract: BACKGROUND:Lumbar spinal fusion in the interbody space is augmented with interbody fusion cages to provide structural support while arthrodesis occurs. Subsidence is a serious complication of interbody fusion. However, the biomechanical influence of anterior longitudinal ligament (ALL) and pedicle screws on subsidence has not been fully understood. OBJECTIVE:To investigate biomechanical effects of the hyperlordotic cages in different surgical conditions using finite element analysis. METHODS:Four surgical finite element (FE) models were constructed by inserting 15 degree lordosis cage at the L3-L4 disc space. The four surgical conditions were ALL intact (M1), ALL resected (M2), ALL intact and bilateral pedicle screws (M3), and ALL resected and bilateral pedicle screws (M4). Follow loads were applied at the L2 vertebral body while the inferior surface of L5 was fixed. FEA was implemented to simulate the four motion modes and biomechanical properties of four fusion scenarios with hyperlordotic interbody cage were compared. RESULTS:The range of motion (ROM) and facet joint force (FJF) at L3-L4 decreased significantly after fusion during all the motion modes. The cage stress and endplate stress at L3-L4 increased significantly after fusion during all the motion modes. The cage stress and endplate stress at L3-L4 for M3 and M4 were smaller than that for M1 and M2 during all the motion modes. The FJF at L3-L4 for M3 and M4 were smaller than that for M1 and M2 during extension, bending, and rotation. CONCLUSIONS:ALL has little effect on the biomechanics after lumbar fusion with hyperlordotic interbody cage. The bilateral pedicle screws significantly decreased the stress in cage, stress in endplate at L3-L4, and lowered facet contact force except for flexion mode. The implication is that the supplemental bilateral pedicle screws are recommended whether or not the ALL is resected.
Keywords: Lumbar spine, biomechanics, lumbar fusion, hyperlordotic interbody cage, finite element analysis (FEA), bilateral pedicle screws, anterior longitudinal ligament (ALL), range of motion (ROM), subsidence, facet contact force
DOI: 10.3233/BME-181004
Journal: Bio-Medical Materials and Engineering, vol. 29, no. 4, pp. 485-497, 2018
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]