Article type: Research Article
Authors: Voice, A.M.; | Vale, M.C. | Fisher, J.
Affiliations: IRC in Polymer Science and Technology, Leeds University, Leeds, LS2 9JT, UK | Medical and Biological Engineering, School of Mechanical Engineering, Leeds University, Leeds, LS2 9JT, UK
Note: [] Corresponding author. Tel.: +44113 233 6647; Fax: +44113 233 3846; E‐mail: [email protected].
Abstract: The depth profile of oxidation index and gel fraction has been measured for two silane crosslinked poly(ethylene) (SXLPE) acetabular cups (one gamma irradiated in air, and one non‐irradiated, both with a shelf‐life of 13 years) and for two UHMWPE components (one gamma irradiated in air and one non‐irradiated, with shelf‐lives of 13 and 7 years, respectively). Only the irradiated UHMWPE exhibited any variation in these properties with depth. The oxidation profile (maximum 1 mm below surface) has been explained to result from reduced levels of diffused oxygen with depth, giving rise to a balance of alkyl and peroxyl radicals (and hence maximum carbonyl production) just below the surface. The gel fraction profile (maximum 4 mm below surface) is also attributed to the lower levels of diffused oxygen with depth, causing crosslinking to dominate in the bulk and chain scission to dominate at the surface. The resistance to oxidative degradation in the non‐irradiated SXLPE was attributed to the use of antioxidants in the polymer processing.
Journal: Bio-Medical Materials and Engineering, vol. 11, no. 1, pp. 37-41, 2001
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]