Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Stuart, J.a; * | Stone, P.C.W.a | Freyburger, G.b | Boisseau, M.R.b | Altman, D.G.c
Affiliations: [a] Department of Haematology, Medical School, University of Birmingham, Birmingham B15 2TJ, UK | [b] INSERM U8, Avenue de Haut-Lévêque, 33600 Pessac, France | [c] Medical Statistics Laboratory, Imperial Cancer Research Fund Laboratories, Lincoln’s Inn Fields, London WC2A 3PX, UK
Correspondence: [*] Correspondence to Professor J. Stuart.
Note: [] Accepted by: Editor G.D.O. Lowe
Abstract: The instrument precision and between-patient variability of haemorheological measurements have been determined using six commercially available instruments: Technicon Ektacytometer, Contraves LS30, SEFAM Erythro-Agregametre, IMH Hemorheometre, Carri-Med St. George’s Filtrometer, and ABX Cell Transit Analyser. Instrument performance was similar at 25°C and 37°C. The different instruments, and sometimes the different parameters of one instrument, showed important differences in precision. Variability of results between patients was often greater than between healthy subjects. The results indicate that both instrument precision and biological variability should be determined in advance of rheological studies so that the required number of subjects can be estimated. A nomogram is presented to facilitate determination of sample size for studies in which the data are to be analysed by either parametric or non-parametric tests.
Keywords: Rheology, Blood viscosity, Erythrocyte deformability, Sample size
DOI: 10.3233/CH-1989-9202
Journal: Clinical Hemorheology and Microcirculation, vol. 9, no. 2, pp. 181-197, 1989
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]