Article type: Research Article
Authors: Tzirakis, Konstantinosa; * | Kontopodis, Nikolaosb | Ioannou, Christos V.b
Affiliations: [a] Department of Mechanical Engineering, School of Engineering, Hellenic Mediterranean University, Heraklion, Crete, Greece | [b] Vascular Surgery Department, Medical School, University of Crete, Heraklion, Crete, Greece
Correspondence: [*] Corresponding author: Konstantinos Tzirakis, Department of Mechanical Engineering, School of Engineering, Hellenic Mediterranean University, Estavromenos, 71410 Heraklion, Crete, Greece. Tel.: +30 2810 379215; E-mail: [email protected].
Abstract: PURPOSE:To compare the hemodynamic performance of three (Bottom Up non-ballet, Top-Down non-ballet, Top Down ballet) idealized stent graft configurations used during endovascular repair of abdominal aortic aneurysms, under the influence of various rheological models. METHODS:Ten rheological models are assumed and a commercial finite volume solver is employed for the simulation of blood flow under realistic boundary conditions. An appropriate mesh convergence study is performed and five hemodynamic variables are computed: the time average wall shear stress (TAWSS), oscillatory shear index (OSI), relative residence time (RRT), endothelial cell activation potential (ECAP) and displacement force (DF) for all three configurations. RESULTS:The choice of blood flow model may affect results, but does not constitute a significant determinant on the overall performance of the assumed stent grafts. On the contrary, stent graft geometry has a major effect. Specifically, the Bottom Up non-ballet type is characterized by the least favorable performance presenting the lowest TAWSS and the highest OSI, RRT and ECAP values. On the other hand, the Top Down ballet type presents hemodynamic advantages yielding the highest TAWSS and lowest OSI, RRT and ECAP average values. Furthermore, the ballet type is characterized by the lowest DF, although differences observed are small and their clinical relevance uncertain. CONCLUSIONS:The effect of the assumed rheological model on the overall performance of the grafts is not significant. It is thus relatively safe to claim that it is the type of stent graft that determines its overall performance rather than the adopted blood flow model.
Keywords: Non-Newtonian blood flow, idealized stent graft, computational fluid dynamics, finite volume method, hemodynamic parameters
DOI: 10.3233/CH-232058
Journal: Clinical Hemorheology and Microcirculation, vol. 87, no. 2, pp. 199-219, 2024
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]