Article type: Research Article
Authors: Kovacs, Davida | Totsimon, Kingaa | Biro, Katalina | Kenyeres, Petera | Juricskay, Istvana | Kesmarky, Gabora | Toth, Kalmana | Toth, Andrasa; b; *
Affiliations: [a] 1st Department of Medicine, Medical School, University of Pecs, Pecs, Hungary | [b] Department of Radiology, Medical School, University of Pecs, Pecs, Hungary
Correspondence: [*] Corresponding author: Andras Toth, MD, PhD, Department of Radiology, Medical School, University of Pecs, 7624 Pecs, Ifjusag ut 13., Hungary. Tel.: +36 72535861; Fax: +36 72536148; E-mail: [email protected].
Abstract: BACKGROUND:Viscosity measurement is challenging due to the internal properties of blood and the artifacts deriving from the various viscometer systems. OBJECTIVE:We aimed to determine the pitfalls of a cone-plate viscometer (Brookfield DV-III Ultra LV) before starting measurements and compare it to our capillary type model (Hemorex Hevimet 40). Effects of sample storage and thermal calibration were assessed as well. METHODS AND RESULTS:Intra-observer variability was studied by 10 replicate measurements of 7 blood samples, mean coefficients of variation were less than 5%. Instruments were compared by measuring 26 blood samples, an average difference of 7% in WBV and 10% in PV was observed. 9 blood samples were stored at 4°C, 22°C and 37°C up to 48 hours to study the effect of storage on viscosity values. WBV at 50 and 100 s-1 became significantly lower after 3 hours at 37°C (p < 0.05). WBV at higher shear rates and PV remained constant at all temperatures. To evaluate the possibility of measuring one sample at different temperatures, 8 blood samples were measured at 40°C with the device calibrated both at 20°C and 40°C; no significant difference was observed. CONCLUSIONS:Thorough validation studies are required before starting experimental and routine viscosity measurements.
Keywords: Capillary, rotational, cone-plate, storage, reproducibility, temperature
DOI: 10.3233/CH-170301
Journal: Clinical Hemorheology and Microcirculation, vol. 69, no. 3, pp. 383-392, 2018
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]