Article type: Research Article
Authors: Khodabandehlou, T.; | Boynard, M. | Guillet, R. | Devehat, C. Le;
Affiliations: Unité de Recherches d'Hémorhéologie Clinique et de Microcirculation, Centre Hospitalier 58000 Nevers, France | Groupe de Recherches en Physique et Biophysique, UFR Biomédicale, Université Paris V, France | Service de Diabétologie‐Endocrinologie‐Maladies Métaboliques, Centre Hospitalier 58000 Nevers, France
Note: [] Corresponding author: Taraneh Khodabandehlou, PhD. E‐mail: urhr@club‐internet.fr.
Abstract: The sensitivity of the ultrasonic interferometry method (Echo‐Cell) to changes in red blood cell (RBC) aggregation was investigated in comparison to the Regulest erythroaggregometry known as a reference method. In experiments where different concentrations of dextrans of 40 or 70 kD molecular weights were added to normal RBCs, the Echo‐Cell was proved as sensitive as erythroaggregometry. A comparative study using RBC samples from normal and diabetic subjects showed that the Echo‐Cell was much more sensitive when aggregation was measured in diluted than undiluted plasma. The sensitivity of Echo‐Cell measurements in diluted plasma was similar to that of erythroaggregometry. Further analysis revealed that RBC aggregation was underestimated by Echo‐Cell when measurements were made in undiluted plasma containing high fibrinogen levels, implying that in that case an elevated plasma viscosity might indirectly affect the sensitivity of the Echo‐Cell. The low sensitivity of the Echo‐Cell to detect an abnormal RBC aggregation when suspensions were prepared in undiluted plasma, is likely related to a relatively high shear stress exerted on RBC aggregates by the suspending medium. In conclusion, the sensitivity of the Echo‐Cell to detect abnormal changes in RBC aggregation can be optimized by diluting the plasma.
Keywords: RBC aggregation, RBC sedimentation rate, dextran, ultrasonic interferometry, Regulest erythroaggregometer, diabetes mellitus
Journal: Clinical Hemorheology and Microcirculation, vol. 27, no. 3-4, pp. 219-232, 2002
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]