Article type: Research Article
Authors: Hardeman, M.R.; | Dobbe, J.G.G. | Ince, C.
Affiliations: Department of Physiology, Academic Medical Center, University of Amsterdam, The Netherlands | Laboratory for Clinical Hemorheology, Academic Medical Center, University of Amsterdam, The Netherlands | Department of Medical Technological Development, Academic Medical Center, University of Amsterdam, The Netherlands
Note: [] Corresponding author: M.R. Hardeman, PhD, Department of Anesthesiology, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands. Tel.: +31 20 566 9111; Fax +31 20 697 9004; E‐mail: [email protected].
Abstract: The Laser‐assisted Optical Rotational Cell Analyzer (LORCA) is unique in its capacity to measure at least two important hemorheological parameters, i.e., deformability of red blood cells (RBC) as well as their aggregation behaviour. In this communication the main principles and characteristics of the aggregometer mode of this instrument are described. Via syllectometry (i.e., laser backscatter versus time), the method allows the measure of both static and kinetic parameters of the aggregation process, e.g., total extent of aggregation, aggregation half time and a combination of these, defined as aggregation index. When the syllectogram deviates from that representing normal rouleaux formation, e.g., in the case of RBC clump formation, this is signalled by a fit error. Maximal flexibility is obtained by various options like an automated re‐iteration procedure (which enables to measure the aggregation tendency and/or the aggregate stability) and aggregation at low shear rate. Experiments dealing with reproducibility, stability and sensitivity of the instrument are described. The latter include a few “classic” methods for inducing subtle changes in RBC aggregation behaviour both regarding cellular factors (heat treatment) and changes in medium constituents (fibrinogen, dextrans). It can be concluded that the aggregometer mode of the LORCA considerably increases the feasibility of this instrument for hemorheological investigations.
Keywords: LORCA, aggregometer, red blood cell, rouleaux formation, red cell aggregation
Journal: Clinical Hemorheology and Microcirculation, vol. 25, no. 1, pp. 1-11, 2001
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]