Article type: Research Article
Authors: Baskurt, Oguz K. | Meiselman, Herbert J.
Affiliations: Koc University School of Medicine, Istanbul, Turkey | Keck School of Medicine, University of Southern California, Los Angeles, CA, USA
Note: [] Corresponding author: Oguz K. Baskurt, MD, PhD, Professor of Physiology, Koc University School of Medicine, Rumelifeneri Yolu, 34450, Sariyer Istanbul, Turkey. Tel.: +90 212 338 1160; Fax: +90 212 338 1168; E-mail: [email protected]
Abstract: Laser-diffraction ektacytometry is a generally accepted technique for measuring RBC deformability induced by fluid shear stress (SS) and yields paired elongation index-SS data at several levels of stress. Unfortunately, comparison of results is hindered by the lack of simple indices that accurately characterize these data. Several mathematical models have been proposed, including those developed for analysis of enzyme kinetics (Lineweaver-Burk, Eadie-Hofstee) and curve fitting (Streekstra-Bronkhorst). All of these analytical approaches provide a value for cell deformation at infinite stress (EImax) and the shear stress required to achieve one-half of this deformation (SS1/2); the use of non-linear regression is essential when calculating these parameters. While the current models provide equivalent results for normal RBC if used with non-linear regression, EImax and SS1/2 are not always concordant for cells with abnormal mechanical behavior. This technical note examines such differences for three conditions: glutaraldehyde treatment, mechanical stress and non-isotonic media. It was found that none of the models yield completely satisfactory values for EImax and SS1/2, especially if there are large changes of EImax. However, the ratio of SS1/2 to EImax (SS1/2/EImax) is much less affected by these problems, has similar power (i.e., standardized difference) as SS1/2 and EImax and is more robust in reflecting alterations of deformability. We thus conclude that the SS1/2/EImax ratio can be used when reporting and comparing various populations of RBC or cells obtained from subjects having different clinical states.
Keywords: Red blood cell deformability, hemorheology, data reduction, ektacytometry
DOI: 10.3233/CH-2012-1616
Journal: Clinical Hemorheology and Microcirculation, vol. 54, no. 1, pp. 99-107, 2013
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]