Issue title: Ergonomic Products
Article type: Research Article
Authors: Lichty, Monica G. | Janowitz, Ira L. | Rempel, David M.;
Affiliations: Battelle Center for Human Performance and Safety, Seattle, WA, USA | Lawrence Berkeley National Laboratory, Division of Environmental Health and Safety, Berkeley, CA, USA | Ergonomics Program, University of California, Berkeley, CA, USA | Division of Occupational and Environmental Medicine, University of California, San Francisco, Richmond, CA, USA
Note: [] Address for correspondence: Monica G. Lichty, Battelle Center for Human Performance and Safety, 1100 Dexter Ave N Ste 400, Seattle, WA 98109, USA. Tel.: +1 206 528 3292; Fax: +1 206 528 3555; E-mail: [email protected]
Abstract: Repetitive pipetting is a task that is associated with work-related musculoskeletal disorders of the hand and arm. Objective: The purpose of this study was to evaluate the usability and ergonomic performance of commercially available pipettes as determined by user ratings and objective measurements. Participants: Participants were laboratory technicians and scientists at the Lawrence Berkeley National Laboratory with experience performing pipetting tasks. Methods: Twenty-one experienced pipette users completed a standardized pipetting task with 5 manual and 5 electronic pipettes. After using each pipette, the user rated it for attributes of comfort and usability. Results: Although no single pipette was rated significantly better than all of the others for every attribute tested, some significant differences were found between pipettes. The Rainin Pipet-Lite received the highest overall quality score among manual pipettes, while the Thermo Scientific Finnpipette Novus was the top-ranked electronic pipette. Features correlated with greater hand and arm comfort were lower tip ejection force, lower blowout force, and pipette balance in the hand. Conclusions: The findings, when considered with participant comments, provide insights into desirable pipette features and emphasize the value of user testing and the importance of the interactions between task, workplace layout, and pipette design.
Keywords: Laboratory, usability testing, tool design, biotechnology
DOI: 10.3233/WOR-2011-1164
Journal: Work, vol. 39, no. 2, pp. 177-185, 2011
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]