You are viewing a javascript disabled version of the site. Please enable Javascript for this site to function properly.
Go to headerGo to navigationGo to searchGo to contentsGo to footer
In content section. Select this link to jump to navigation

Erratum to: The vestibular implant: Opinion statement on implantation criteria for research

[Journal of Vestibular Research, 30(3) (2020), pp. 213-223. DOI: 10.3233/VES-200701]

On page 215, under section 3, “VI-implantation criteria”, where it reads:

Reduced horizontal angular VOR gain≤0.1 upon sinusoidal stimulation on a rotatory chair (0.1 Hz, Vmax = 50°/sec) and a phase lead > 68° (time constant < 5 sec)

Should be changed to:

Reduced horizontal angular VOR gain≤0.1 upon sinusoidal stimulation on a rotatory chair (0.1 Hz, Vmax = 50°/sec) and a phase lead≥15° (time constant≤6 sec)