Article type: Research Article
Authors: Ulivieri, Fabio M. | Piodi, Luca P. | Vezzoli, Michela | Verdoia, Cesare | Gerundini, Paolo | Pedotti, Paola | Taioli, Emanuela; *
Affiliations: [a] Ospedale Maggiore di Milano IRCCS, Milano, Italy | [b] Istituto Ortopedico G.Pini, Università di Milano, Italy
Correspondence: [*] Address for correspondence: Emanuela Taioli, MD PhD, Ospedale Maggiore di Milano IRCCS, Padiglione Marangoni – Molecular and Genetic Epidemiology Unit, Via F.Sforza 35, 20122 Milano, Italy. Tel./Fax: +39 0255034055; E-mail: [email protected]
Abstract: To determine the reliability and validity of a new ultrasound bone densitometry instrument, we studied 41 healthy volunteers (27 females and 14 males, mean age 33.8 yrs, range 22–45 yrs), who underwent ultrasound measurements with QUS-2 (Metra Biosystems Instrument [Quidel]). The measurements were repeated three times in the same visit (mean coefficient of variation 3.4%, range 0.64–10.87%). Fifteen of these subjects repeated the three measurements in two subsequent visits, 7 days apart from each other. There was a significant variability between subjects (p = 0.0001) and within subjects (p = 0.004). To compare the measurement performed with the new equipment with that of the gold standard, 19 subjects underwent a bone mass measurement by dual X-ray photon absorptiometry (DXA, Hologic QDR 4500A). The broadband ultrasound attenuation correlated with the total femur and femoral neck bone mineral density (r = 0.61, p = 0.005; r = 0.4, p = 0.09, respectively), but not with lumbar spine r = 0.34, p = 0.15). The mean coefficient of variation of the three ultrasound measurements performed in each of the subjects was 3.4% (range 0.64–10.87%). QUS-2 Metra Biosystems ultrasound densitometer does not seem at the moment a useful tool to assess bone mass in clinical setting, because of its high coefficient of variation and its high intra- subject variability.
Keywords: QUS-2 Metra Biosystems, ultrasound densitometer, broadband ultrasound attenuation (BUA), osteoporosis
DOI: 10.3233/BMR-2004-17201
Journal: Journal of Back and Musculoskeletal Rehabilitation, vol. 17, no. 2, pp. 37-40, 2004
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]