Article type: Research Article
Authors: Froelich, Jens J.a; * | Ray, Udayanb | Monkhorst, Jessicaa | Marwick, Thomas H.c | Hardikar, Ashutoshd | Harle, Robina | Carr, Michael W.a
Affiliations: [a] Department of Medical Imaging, Royal Hobart Hospital, Hobart, TAS, Australia | [b] Department of Pathology, Royal Hobart Hospital, Hobart, Australia | [c] Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia | [d] Department of Cardiothoracic Surgery, Royal Hobart Hospital, Hobart, Australia
Correspondence: [*] Address for correspondence: Jens J. Froelich, Prof. Dr., Department of Medical Imaging, Royal Hobart Hospital, 48 Liverpool St., Hobart, TAS, 7000, Australia. Tel.: +61 3 6222 8612; Fax: +61 3 6222 8061; E-mail: [email protected].
Abstract: Background:Microcatheter directed blood reperfusion is an endovascular salvage option for acute cerebral artery occlusions. It has not been investigated whether this technique may be associated with hemolysis. Objective:Analysis of hemolysis during blood infusion through different microcatheters and infusion rates to assess related risks. Methods:Four microcatheters with different inner diameters were perfused with blood samples at three infusion rates. Hemolytic markers including lactate-dehydrogenase (LDH) and haptoglobin were analyzed. Samples before and after blood infusion were compared using Student’s t-test. Flow-related degree of hemolysis was analyzed with regression analysis. Resulting shear stress was calculated and correlated with LDH and haptoglobin. Results:Significant increase of LDH and decrease of haptoglobin was found after blood reperfusion through small microcatheters at progressive flow rates (p<0.05). No hemolysis was found with larger diameter microcatheters at all flow rates (p>0.05). Correlation between shear stress, LDH and haptoglobin was r=0.86 and r=0.75, respectively. Conclusions:Progressive hemolysis occurs during blood perfusion of small lumen microcatheters at increasing flow rates. This phenomenon may be related to turbulent flow, exposure time and increased shear stress. Larger microcatheters did not induce hemolysis and may be the preferred choice for stroke reperfusion.
Keywords: Stroke, ischemia, hemolysis, catheterization, blood, reperfusion
DOI: 10.3233/BIR-15074
Journal: Biorheology, vol. 52, no. 4, pp. 279-291, 2015
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]