Issue title: The Fifth Conference of the International CoQ10 Association, Kobe 2007 – 50th anniversary of CoQ10 discovery
Article type: Research Article
Authors: Bompadre, Stefano | Tulipani, Sara | Romandini, Stefania | Giorgetti, Raffaele | Battino, Maurizio
Affiliations: Institute of Microbiology and Biomedical Sciences, Faculty of Medicine, Università Politecnica delle Marche, Ancona, Italy | Institute of Biochemistry, Faculty of Medicine, Università Politecnica delle Marche, Ancona, Italy | Institute of Legal Medicine, Faculty of Medicine, Università Politecnica delle Marche, Ancona, Italy
Note: [] Address for correspondence: Dr. Maurizio Battino, PhD, DSc, MD Hon, Inst. of Biochemistry, Faculty of Medicine, Università Politecnica delle Marche, Via Ranieri, 65, 60100 Ancona, Italy. Tel.: +39 071 2204646; Fax: +39 071 2204398; E-mail: [email protected]
Abstract: A novel isocratic modified column-switching HPLC method for automated quantitative analysis of Vitamin E and Coenzyme Q, in the reduced and oxidized form, is described. Many column-switching HPLC methods are found in the literature, also for determining antioxidant substances, but we developed a different system of column-switching. An empty column, 5 cm long, was connected to the switching valve, before the sample loop and the extraction column. The sample loop was connected directly after the empty column. The inserted column, containing about 1.4 ml of the extraction eluent simulated a gradient elution, enhancing sensitivity and resolution. When switching the columns, the empty column is placed right before the extraction column and acts as a static mixer for the extraction phase and the incoming analytical phase. Samples were cleaned from interfering compounds by transfer onto a extraction-column, using a C-8 silica. Separation was performed onto an analytical column C18 3 μm, 150 mm × 4.6 mm at a flow rate of 1.0 ml/min with 20 mmol/l lithium perchlorate/perchloric acid, pH3.0 in Ethanol as analytical eluent. Detection was performed with a ESA Coulochem 5100 A model. The method was found to be suitable for automated analysis of Coenzyme Q, reduced and oxidized form, and Vitamin E in serum.
Keywords: Solid-phase extraction, column-switching, HPLC-EC
Journal: BioFactors, vol. 32, no. 1-4, pp. 257-262, 2008
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]