Article type: Research Article
Authors: Myers, Edwina; * | Triscari, Robertb
Affiliations: [a] Occupational Therapy Program, Department of Rehabilitation Sciences, Marieb College of Health and Human Services, Florida Gulf Coast University, Fort Myers, FL, USA | [b] College of Education, Florida Gulf Coast University, Fort Myers, FL, USA
Correspondence: [*] Address for correspondence: Edwin Myers, Department of Rehabilitation Sciences, College of Health Professions and Social Work, Florida Gulf Coast University, 10501 FGCU Boulevard South, Fort Myers, FL 33965, USA. Tel.: +1 239 590 7551; Fax: +1 239 590 7460; E-mail: [email protected].
Abstract: BACKGROUND: The purpose of this study was to provide evidence regarding the Baltimore Therapeutic Equipment (BTE) Work Simulator II’s recommended grip endurance protocol. The grip endurance protocol of the BTE Simulator II has never been validated, though it has been used often for the rehabilitation of work-related injuries and other occupational dysfunctions. Without validation, the grip endurance protocol may or may not be providing skilled clinicians with appropriate evaluation results. This study evaluated a protocol comparing the BTE Simulator II to the Jamar Dynamometer to determine how the BTE compares to the Jamar device. OBJECTIVE: To establish whether the recommended grip endurance protocol for the BTE Simulator II is comparable to the parameters established for the Jamar handgrip dynamometer. METHODS: Data were gathered from 140 participants ages 18 to 40 at the time of the study. Participants completed protocols established for the BTE Simulator II and compared to a performance of a grip strength endurance protocol for the Jamar Handgrip Dynamometer. After establishing baseline strength levels for the participants, they were timed to see how long they could sustain a 30% maximum voluntary contraction (MVC) on each device. Sample t-tests were calculated to compare the results of the BTE Simulator II to the previously validated Jamar device. RESULTS: A dependent sample t-test found no statistically significant difference between the times a participant sustained 30% of their maximum voluntary contraction (MVC) on the BTE Simulator II with attachment #162 versus the Jamar Dynamometer when comparing the differences of the means. This provides possible evidence of validity for the BTE endurance protocol. The independent sample t-test found no statistically significant difference between the grip endurance of the males versus the females, however, the means indicate men sustained 30% of their MVC longer than their female counterparts on the BTE. CONCLUSIONS: The findings indicate that the 30% MVC may be a reliable baseline for grip strength endurance testing following the BTE Simulator II protocol. This suggests that the BTE Simulator II’s endurance protocol may be a useful tool to document client progression during rehabilitation after sustaining an upper extremity dysfunction. Further research is needed to validate this protocol using different populations.
Keywords: Muscular endurance, functional capacity evaluation, grip endurance, maximum voluntary contraction (MVC), grip strength
DOI: 10.3233/WOR-172542
Journal: Work, vol. 57, no. 1, pp. 95-103, 2017
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]