Instrumental analysis applied to erythrocyte properties

Abstract

The properties, structure and functions of the erythrocyte or red blood cell (RBC) have been known and evaluated by using simpler to more sophisticated technical devices. Microscopy, flow cytometry, spectrometry, spectroflurometry, aggregometer, ektacytometer zeta potential, amperometry, electrophoresis and centrifugation, are apparatus used to qualitatively and quantitatively characterize RBC, based on the principles of light transmission, light scattering spectroscopy, light absorption, fluorescence, light polarization, shear stress, shear rate, charge and molecular weight gradients. There is a symbiosis between the chemical, the physical and the mechanical principles of those instrumental analysis and the RBC properties to be measured. Here we briefly exemplify the relationship between the biochemical, biophysical and mechanical based techniques and the properties of the erythrocyte in healthy and disease.

1Erythrocyte bioavailability in oxygen and nitric oxide

Since the antiquity blood had fascinated the humans who associated it to curative power and magic process. In the seventeen century Jan Jacob Swammerdan (1637–1680) was the first scientist to observe erythrocytes or red blood cells (RBC) from frog blood under the light microscope. He was followed almost in same time by Antonje Van Leeuwenkoek (1633–1723) who performed the RBC observations also with light microscopy but equipped with lenses produced by himself (in http://micro.magnet.fsu.edu/primer/museum/swammerdam1670s.htm and http://pt.wikipedia.org/wiki/Anton_van_Leeuwenhoek). He observed that the erythrocyte can elongate up to three times their original dimension, can undergo aggregation and he estimated the diameter of RBC as 8.5μm. The erythrocyte is a component of blood and its ability to delivery oxygen to all tissues of the vertebrates, makes it unique. The red color of the erythrocytes raises the curiosity in those scientists to look them through the light microscope. Hemoglobin, that is the carrier protein of oxygen from lungs to tissues, absorbs red and infrared light differently [1].

The percentage of haemoglobin saturated with oxygen can be measured in real time by pulse oximetry and is an indicator of the blood oxygen saturation [2]. Oximetry consist in a probe that can be attached to the patient‘s finger, nose, toe or earlobe lied to a computerized monitor which allows a visual waveform and an audible signal emitted with each pulse beat (the tone decreases with decreasing saturation). It is a non-invasive device to monitor the hemoglobin oxygen saturation (SaO2) of arterial capillary blood that uses a light source and a light detector. Desaturated haemoglobin absorbs red light and oxyhemoglobin infrared light. If the oximeter measures greatest absorbance in the red band it will indicate low saturation. If the higher absorbance is in the infrared band it means high saturation usually related with blood filled in the arteries. It needs attention to several interfering factors like exposition of patients to carbon monoxide or a situation of cardiac arrest [2]. For estimation of the tissue oxygenation status of patients it is necessary to quantify the partial arterial oxygen pressure (PaO2) and P50 assessments. The affinity of haemoglobin to oxygen can be described as the oxygen tension at which the haemoglobin is 50% saturated (P50). The value of P50 indicates the possibility of oxyhemoglobin deliver oxygen to tissue and can be evaluated mathematically applying the Hill equation developed by Lichtman et al. [3]. There is an instrument where the oxygen tension is detected by a Clark electrode and the oxyhemoglobin fraction (% HbO2) is evaluated by a dual-wavelength spectrophotometer and the dissociation curve (a relationship between the amount of oxygen dissolved in the blood and that is attached to the haemoglobin) is registered in a scanner [4]. High and low values of P50 indicate decreased and increased affinity of oxygen to haemoglobin respectively. The intra globular increased content of certain allosteric molecules, like 2,3 bisphosphoglycerate carbon dioxide, and the protons (decrease of the pH values) increased P50 value [5, 6]. Using circular dichroism spectroscopy large functional changes can take place, at molecular levels, between hemoglobin in R state (ligated with oxygen) and those heterotropic ligands that decrease the hemoglobin affinity for oxygen even in a full ligated state [7]. Changes in SaO2, PaO2 and P50 values were obtained when determined under physiological conditions like exposition to high altitude or during physical exercise [8]. At high altitude atmospheric pressure decreases, then PaO2 decreases and consequently low haemoglobin oxygenation occurs in non adapted healthy persons that start to manifest dyspnoea, tachycardia, tiredness or more complicated events. In adapted humans such as the highlanders those adverse issues are not found. They have higher values of P50 (30 mmHb versus normal 27 mmHg) [8].

During maximal exercise the amount of lactate increases, acidosis can occur in muscle, liver and kidney, blood flow can be compromised leading to a decreased tissue oxygen tension [8a]; the experimental measurements showed left shift in the oxyhemoglobin dissociation curve, lower P50 value (increase the haemoglobin affinity to oxygen) which reflect decreases delivery of oxygen [8].

Mutations on haemoglobin gene like in sickle cells present high values of P50 and lower PaO2 given by pulse oximetry [9]. Low haemoglobin affinity for oxygen and low oxygen saturation in peripheral blood was also verified by pulse oxymetry in patients presented mutation on α-globin chain, named haemoglobin Titusville [10].

It was found that the P50 was increased in hypertensive patients with diastolic pressure higher or equal to 130 mm Hg [11]. In a group of diabetic patients with different degrees of retinopathy normal levels of erythrocytic 2,3BPG and normal enzyme activities of the hexokinase, phosphofrutokinase and pyruvate kinase were obtained but in spite of these they presented high plasma glucose levels and decrease values of P50 [12]. This higher haemoglobin affinity for oxygen suggested transient decreases in RBC oxygen delivery with implications on retinal microvasculature structure [12]. At this time the authors proposed the influence of the glycosylated haemoglobin (HbA1c) concentration in the lower value of P50 as an explanation [12] which was not corroborated later [13–15]. A spectro-fotometric method was initially described [16] for quantification of HbA1c and was further modified by us [17]. Nowadays there are commercial diagnostic kits for quantification of glycosylated hemoglobin allowing diabetic patients to have it in their routine analysis. For clinicians the value of HbA1c serves as a marker of average blood glucose levels over the last 2-3 months.

It was shown that the P50 values are gender-independent and increased when adrenaline was added to blood samples obtained from healthy adult female and male [18]. In other “in vitro” studies after incubation of blood samples in presence of acetylcholine an increase of P50 values was verified demonstrating a decreased affinity of haemoglobin to oxygen [19]. An exogenous molecule like Spermine NONOate, a NO donor, was shown to increase the hemoglobin affinity for oxygen [19].

It must be noted that higher levels of hemoglobin oxygenation stimulate the anion exchange (band 3 protein) while deoxygenation accelerates the potassium-chloride co-transport and decrease the mechanical fluctuations of the membrane (CMFs) [20, 21]. The CMFs may be monitored by time-dependent light scattering from a small area of the cell surface by a method based on point dark field microscopy [21]. The mechanical fluctuations are associated with membrane displacement devoid to a contribution of hemoglobin, to thermal and metabolic energies and to erythrocyte deformability [21].

Acceleration of glycolysis (glycolytic enzymes leave the band 3 protein to cytosol), occurs concomitant with binding of deoxyhemoglobin to band 3 protein [22].

Beyond the ability of haemoglobin to deliver “in vivo oxygen” other factors related with the tissues ability to extract it and with microvascular haemodynamics are determinants for tissue oxygenation. The use at bed-side of sub-lingual device, like sidestream dark field imaging (SDF), allows the evaluation of blood flow characteristics [23]. SDF uses a polarized light to see and measure the number of small vessels either in occlusion or in perfusion [24]. For blood flow evaluation in the human tympanic membrane laser-Doppler flowmetry was used and ameliorated recently with a vibrometer approach [25, 26]. The identification of impaired microvascular blood flow by those noninvasive means can lead to early identification of patients at risk for peripheral or vascular and coronary artery disease [27, 28].

Nitric oxide was identified as a vasodilator in dependence of the intactness of the vascular endothelium in presence of acethylcholine [29]. Erythrocytes scavenge NO from endothelium bound to the heme of deoxygenate hemoglobin forming nitrosyl haemoglobin and tied to the cysteine of oxyhemoglobin beta chain originating S-nitrosohemoglobin [30, 31]. Electron paramagnetic resonance and photolysis-chemiluminescence techniques were used to quantify S-nitroso and nitrosylhemoglobin [30, 31]. Cristallographic analyses confirmed the binding of NO to the heme group of T state (deoxygenated) of hemoglobin while the NO binding to beta cysteine needs aerobic conditions [32].

The rate of NO influx into erythrocytes is greater when haemoglobin is oxygenated and in lower hematocrit than when is oxygenated [33]. The influx is blocked by Heinz body formation or by inhibitors of anion exchange also named band 3 protein [34].

When a situation of higher tissue oxygen demand exist a signal transduction mechanism occur involving cAMP with efflux, from erythrocyte, of adenosine triphosphaste (ATP) which is recognized by the purigenic receptor in endothelial cell and consequently NO synthesis is stimulated initiating vasodilation [35]. In situations of hyperemia the nitrosylhemoglobin values measured by electron paramagnetic resonance correlated with the tonometry data’s of the endothelial function [36].

The visualization of nitric oxide inside erythrocytes was obtained with fluorescence microscopy digital images by loading with diaminofluoreceine-2diacetato (DAF-2D) that in presence of NO, oxygen and esterases originate triazolofluorescein [37]. The synthesis of NO with the participation of nitric oxide synthase (NOS) inside erythrocytes has been described using immunogold-labeling and electron microscopy imaging, immunofluorescence confocal microscopy, Western blotting, and reverse transcriptase – polymerase chain reaction [38]; RBC NOS present structural similarities with endothelial source [38]. Others used immunohystochemistry evaluated semi-quantitative immunoreactivity of the 1177 phosphorylated residue of NOS protein [39]. Other authors evidenced their doubts by gas chromatography-mass spectrometry assays about the erythrocyte enzyme activity able to generate sufficient effective NO [40].

The NO effluxes from erythrocytes can me measured by an amperometric method [41]. NO diffuses through the gas-permeable membrane triple COAT of the sensor probe and is then oxidized at the working platinum electrode, resulting in an electric current. The redox current is proportional to the NO concentration outside the membrane and is continuously monitored with the electrochemical detection system and connected to a computer [41]. For more details see the review about the application of NO sensors im medicine [42]. A higher NO release from erythrocyte samples obtained from patients with diseases related with hypoxia and inflammatory states, namely sickle cell disease, hypercholesterolemic and hypertensive patients was verified, besides the fact that they sustain impaired erythrocyte deformability [43]. Glutathione is another important molecule that binds NO forming S-nitrosoglutatione (GSNO) which amounts inside the erythrocytes was quantified by a spectrophotometric method [44]. However using this method the efflux of GSNO from erythrocytes was only verified when erythrocytes were stimulated with acetylcholine and or timolol maleate [45].

2Erythrocyte membrane properties and functions

The qualitative and quantitative observation of RBC membrane properties and functions highlight some of the blood flow behaviour in the vascular network. In the end of ninety century Ehrlich and Romanowsky used not successfully different aqueous dyes to stain blood smears and William Leisham and James Wright started to use methanol to fix the cells before staining to surpass that difficulty [46].

The shape of RBC can be observed in the optic microscope with limiting resolution of 0.2 microm and absence of the third dimension (thickness). The appearance of the scanning electron microscope strongly ameliorates those limitations allowing the visualization in 3D with tenfold of improvement in resolution [47]. Since the seventeen’s electron microscope progress to eliminate the metal or carbon coating around the samples to be capable of “low vacuum” operation [48–50].

Besides its property to magnify cells it does not show the changing movements that characterize a living cell.

In 1953 the Nobel Prize in physics award Frederik Zemike for the development of the phase contrast microscope which was acquired by many basic science laboratory [51].

It continues facing the assessment of qualitative analyses without subcellular specificity that means do not distinguish below half the wavelength of light although the high magnification attained.

At same time the appearance of the fluorescence microscopy and of the confocal microscope brought new possibilities to study erythrocyte properties and binding to molecules. Now we are able to have molecular specificity and high spatial resolution but with loss of live cells in their native physiological state due the requirement of cell permeabilization and immunostaining [51a].

In October of 2014 the Nobel Price in Chemistry went to Eric Betzig, William Moerner and Stefan Hell for their development of a super-resolved fluorescence microscopy which brings optical microscopy into the nanodimension [52].

Using fluorescence confocal microscopy and flow cytometry it was possible to show that erythrocyte binds soluble fibrinogen in dependence of age in a way that the younger subpopulation of erythrocytes binds much more fibrinogen molecules than the older of the same blood sample [53]. All subpopulations, the older and the younger erythrocytes decrease its ability to bind fibrinogen when antibody CD47 was added previously. Noticed that CD47 is a membrane protein present in different cells with the purpose to signalize them to be not “to eat” by phagocytes and consequently advertized notto be removed from blood circulation. Using zeta potential technique that measure the extension of interaction between two particles it was able to confirm the binding between erythrocyte CD47 and soluble fibrinogen in dependence of the age of RBC subpopulations [53]. For this study the erythrocytes were obtained from centrifugation of blood samples and are separated in subpopulations of different ages by percoll discontinuos gradient as described by Corsi et al. [54] and Venerando et al. [55] with some modifications [53].

Using phtalates esters mixtures of different gradients it is possible to isolate exovesicles produced from erythrocyte when submitted to fluorescent probes (usually used for membrane fluidity measurements) that were not visible in buffer suspensions [56]. In order to visualize them coloration for protein (Commassie blue staining), for acetylcholinestease (Ellman‘s reagent for the enzyme spectrophotometric assay) and for phospholipid content (phosphatidic acid coloration) were used [57, 58]. The importance of these coloration processes is to avoid the physically evaluation of exovesicles by using light scattering spectroscopy that is most expensive.

The human RBC in blood circulation has a prolonged life-span of 120 days and present susceptibility to aging and eryptosis (apoptosis) which produce exovesicles liberation [59, 60]. Acetylcholinesterase (AChE) is a biomarker of exovesicles originated from erythrocytes [56, 61] and also is considered as an index of RBC membrane integrity.

Increase enzyme activity of erythrocyte AChE has been reported in arterial hypertension, lateral sclerosis amyotrophic, and deficient in paroxysmal nocturnal hemoglobinuria [62–64]. Erythrocyte AChE is a very peculiar enzyme which is inhibited by high concentrations of acetylcholine (ACh) its natural substrate [65]. So, AChE originates active and inactive and less active complex forms of the enzyme according its association with the substrate (in concentration below Km) or with stronger inhibitors respectively [65].

The enzyme complex forms formed in the erythrocyte membrane between AChE and ACh or AChE and velnacrine (strong inhibitor) participate in the nitric oxide transduction mechanism associated with the band 3 protein phosphorylation/dephosphorylation through the Gi protein [66, 67]. In these studies, spectrophotometry, electophoresis and immunoblotting techniques were used and it was evidenced that AChE function and a chemical transducer [66, 67]. It was shown that the erythrocyte in the presence of acetylcholine increased its ability to delivery NO which disappears if fibrinogen was added [68]. Fibrinogen known as an acute phase protein maintains the NO inside the erythrocyte as shown in anin vitro study [69]. In hyperfibrinogenemia the scavenger ability of the erythrocyte to maintain NO inside prevails also in presence of the 4N1K the agonist of CD47 with increased GSNO, peroxinitrite, nitrite and nitrate concentrations [70]. Otherwise high fibrinogen levels concomitant with band 3 phosphorylation promote the efflux of NO from the RBC [71].

Phosphorylation of band 3 protein promotes the (i) displacement of the glycolytic enzymes to the cytoplasm from its N terminus and the approach of S – nitrosohemoglobin (ii) increase of the efflux of NO from erythrocyte, (iii) enzyme activity of AChE and (iv) the erythrocyte aggregation [22, 66, 67, 72, 73].

Erythrocyte aggregation is measured using either light transmittance or light back-scattering following abrupt stop of shear through or from red blood cell suspensions (anticoagulant blood or in autologous plasma) respectively [74]. Both measure the time course and the intensity of the aggregation but a faster time course was obtained with the light back-scattering if compared to the value resulting from the measurement done with light transmittance [74]. After shearing the blood sample the apparatus stop and the laser backscatter intensity from the samples versus time is recorded (syllectogram) before and after sudden stop (zero time). This method failed for blood samples taken from patients with very strong RBC aggregates. The use of ultrasound with an echo probe substitute backscatter light and allow in vivo evaluations [75].

In the rotative cone-plate aggregometer, that disperses the blood cells by high shear stress (600/s) connected to a photometer, a syllectogram is also obtain (intensity of infra red light transmission versus time) that determines the extent of aggregation by integrating for 5 or 10 seconds [76, 77]. An update was performed connecting a computer to provide kinetic data of aggregation process [78]. Another apparatus was developed with shear rates from zero to 3500/s and measurements are based on changes in laser backscatter intensity [79].

Beyond those methods there are others based on (i) microscopy with computerized image analysis, (ii) low shear viscometry and electrical properties are also used and can be reviewed in the Handbook of Hemorheology and Hemodynamics [80].

Fibrinogen, lipoproteins, macroglobulins or immunoglobulins, plasma viscosity and hematocrit affect erythrocyte aggregation [81]. Plasma viscosity is an extrinsic influent factor in erythrocyte aggregation and if it is excluded by suspending RBC in standardized suspending media for example isotonic dextran solutions the greater or lower tendency to aggregate is called aggregability [82, 83]. The tendency of RBCs to reversibly aggregate and disaggregate influences blood flow [84].

RBC aggregation interfere with the measure of the erythrocyte sedimentation rate (ESR) assessed in routine clinical laboratories because it is recognized as a marker of acute phase of inflammation [85]. ESR is determined in vertically mounted tubes of defined length and diameter after one hour as first described by Westergren in 1921 [86]. Using blood samples with anticoagulant (citrate) the drop of RBC under the gravity force is also influenced by its size and shape as well the values of plasma viscosity, fibrinogen levels, hematocrit and room temperature where the test is done. Based on this method other was developed to spend only 20 min in the RBC drop by scanning twice the tube containing the anticoagulant blood sample with an optoelectronic light source [87]. A shorter assessment of ESR in 20 seconds was created using a microflow cell in a photometric rheoscope but the ESR values obtained were lower correlated with those assessed by the reference method of Westergren [88].

High erythrocyte aggregation values were shown in blood samples obtained from hypertensive patients besides the decreased values of plasma viscosity and hematocrit which evidenced the contribution of the intrinsic properties of RBCs on its tendency to aggregate [89–91]. The haemoglobin oxygen content did not cause interference in erythrocyte aggregation if measured by light transmittance [92].

Erythrocyte aggregation and plasma viscosity were elevated in patients with different degree of retinal vascular damage which are associated with plasma viscosity [93].

The association between the strength of large RBCs aggregates and the high plasma fibrinogen concentration was observed in patients with inflammatory bowel disease with detrimental microcirculatory blood flow in the intestinal microvasculature [94]. RBC aggregation was reported to be a biomarker of cardiovascular disease due to its association with other cardiovascular risk factors as observed in obesity, hypertension, dislipidemia and diabetes [95–98].

A review about the erythrocyte aggregation values obtained in patients monitorized by cross-sectional and longitudinal clinical studies was published [99].

Erythrocyte aggregation and deformability determine blood viscosity and microvascular perfusion [100, 101, 101a]. While erythrocyte deformability influences both blood flow in macro- and microcirculation, erythrocyte aggregation affected it predominantly in low-shear regions of microcirculation [102]. RBC when exposed to high shear stress and forced to pass through capillaries that can be as small as one quarter of the cell diameter are able to reversibly deform (change reversibly its form without alteration of the volume) which means present deformability [103].

The partially or totally aspirated erythrocyte into capillaries of 1 to 5 microns was quantified based on the amount of negative pressure yielding the shear elastic modulus of the membrane or the deformability [104]. It was a method not widespread due to its time consuming requiring especial skills. Several years have been passed to develop a microfluidic approach that provides morphological information at the single-cell level [105]. It was evidenced with this single-cell technique no changes in erythrocyte deformability when storage bank blood samples were assessed [105].

The first home technique to measure the ability of RBC to deform was developed by Reid et al. [106]. It is a filtration method based on gravity force or by application of a positive or negative pressure. The passage of a certain amount of erythrocyte suspensions (Ht 8% ) through 3 to 5 microns pores was quantified by time taken in relation to buffer or by pressure-flow relationship [106]. Using the Reid et al. method in blood samples obtained from diabetic patients with microangiopathy a decreased filtration rate was verified [12]. Patients suffering from cerebral disorders presented also a decrease of the deformability index in a study which the Reid et al. apparatus [107]. Trying to simulate capillaries another device was constructed with flow channels connected to a microcomputer. This computer–controlled filtrometer measured the flow curve of a RBC suspension trough a filter membrane [108]. The apparatus contained U-shaped tubes and RBC suspension is place in one side and the filter is inserting at the collateral side of the U-shaped tube [108, 109].

A microchannel coupled with a microscope is named a rheoscope setup useful to visualize the RBC while being deformed instead to obtain mean of deformability index; it is available to follow shape changes under a specific shear stress [110–112].

Erythrocyte deformability can be determined using the Rheodyn SSD laser diffraction ektacytometer (or ellipsometry) that was initially developed by Goner et al. [113]. The Rheodyn SSD diffractometer determines erythrocyte deformability by simulating the shear stresses exerted by the blood flow and vascular walls on the erythrocytes. Erythrocytes are suspended in a high viscous medium and placed between a rotating optical disk and a stationary disk, where they are going to be subjected to well defined shear stress, which forces the erythrocytes to deform to ellipsoids and align with the fluid shear stress. A computer incorporated gives the analysis of the different ellipse, curves according the shear stress, and elongation index (EEI) are obtained respectively. The respective curve profile resulting from all EEI values in function of the shear stress applied can be linearized as a Lineweaver-Burk analysis procedure.

Beyond erythrocyte membrane composition and structural arrangement of its macromolecules, the internal viscosity and the ratio between RBC volume and its surface are important parameters on erythrocyte deformability [114].

Computer simulations compared with in vitro results were recently studied and show that the characteristic transient time depends on the imposed flow strength in RBC passing through small microcapillaries [115]. The numerical simulated data are validated by the theoretical analysis of experimental data which give the values of RBC 2D membrane viscosity and confirms the characteristic transient time value with those obtained by other technical devices [115].

Lower oxygenation of haemoglobin impaired the erythrocyte deformability [116]. PKC inhibitors also decreased RBC deformability [117]. It was evidenced that elimination of phosphorylation of protein 4.1 allows stronger interactions between membrane proteins with those of cytoskeleton [118].

Erythrocte deformability is improved in vitro by sodium nitropruside, an NO donor, as much as the deoxygenate state of haemoglobin [116]. Erythrocyte under shear stress by ektacytometer in presence of internal or external stimuli showed different erythrocyte deformability curves [119].

Carbenoxolona, an inhibitor of pannexin-1 hemichanel exporter of ATP, increased erythrocyte deformability in lower shear stress evaluated by ektacytometer [120].

Impaired erythrocyte deformability-dependence of plasma glucose level has been reported in diabetes with coronary artery disease that can increase morbidity in these patients [121]. Patients with nephropathy with or without diabetes presented decreased erythrocyte deformability assessed byektacytometer[122].

Usually humans have their routine blood tests according their health system and physiological state. For that, clinical laboratory of analysis use more or less sophisticated Coulter counter instruments based on the detection of electrical conductance changes resulting from the passage of RBC in a conductive medium [123]. For example the RBC distribution width (RDW) is an important characteristic of the performance in athletes and present prognostic value in cardiovascular pathologies [124, 125].

An approach to measure simultaneously RBC area and volume under flow in microcapillaries, using high-speed video microscopy imaging was developed and data compared with Coulter counter and micropipette assays [126]. This technique overcomes the reliability of micropipette experiments, eliminates the conductivity of the suspension medium and assures the measurements of RBC surface area, (omitted in the routine Coulter count), volume, RDW and reproducibility of the results [127].

3Conclusions

The oxygenation or deoxygenation status of hemoglobin depends on a myriad of properties inherent (i) to RBC (shape, membrane stability, metabolism, oxidative stress and hemoglobin structure and binding to nitric oxide and to allosteric modulators), (ii) to blood characteristics namely its composition, type of flow, rheology, (iii) to vascular environment related to hemodynamic properties, hemostatic processes, inflammation status, presence of atherosclerosis plaque and (iv) to tissues capacity and metabolic needs to extract oxygen.

The hemoglobin oxygenation level critically influences intracellular signaling pathways, action of hormones and or vasoactive agents, ion transport, and deformability of RBC.

The erythrocyte bioavailability in nitric oxide (NO) or its tendency to deliver or scavenge it depends on (i) endogenous and exogenous compounds and (ii) erythrocyte redox status and membrane protein phosphorylation degree; and affects erythrocyte deformability. All these erythrocyte properties and functions have been discovered among the years with the aid of instruments based on biochemical, biophysical and biomechanical characteristics of RBC. Computer simulations were also developed but were beyond the objective of the present review.

References

1 

Perutz MF1978Hemoglobin structure and respiratory transportSci Am23992125

2 

Grap MJ1998Pulse oximetryCrit Care Nurse189499

3 

Lichtman MA, Murphy M, Pogal M1976The use of a single venous blood sample to asses oxygen binding to haemoglobinBrit J Haematol328998

4 

Guarnone R, Centenara E, Barosi G1995Performance characteristics of hemox-analyser or assessment of the hemoglobin dissociation curveHaematologica80426430

5 

Nikinmaa M1980Vertebrate Red Blood CellsSpringerBerlin

6 

MacDonald R1977Red cell 2,3-diphosphoglycerate and oxygen affinityAnaesthesia32544553

7 

Tsuneshige A, Park S, Yonetani T2002Heterotropic effectors control the hemoglobin function by interacting with its T and R s T tates–a new view on the principle of allosteryBiophys Chem984963

8 

Nielsen HB2003Arterial desaturation during exercise in man: Implication for O2 uptake and work capacityScand J Med Sci Sports13339358

{ label needed for ref[@id='r8a'] } 

Jung F, Keßler H, Pindur G, Sternitzky R, Franke RP1999Intramuscular oxygen partial pressure in the healthy during exerciseClin Hemorheol Microcirc212533

9 

Abdu AL, Gómez-Márquez J, Aldrich TK2008The oxygen affinity of sickle haemoglobinRespir Physiol Neurobiol1619294

10 

Luo H, Irving I, Prior J, Lim E, Eung SH, Skelton TP, Erber WN, Steinberg MH, Chui DHK2004Hemoglobin Titusville a low oxygen affinity variant hemoglobin in a family of Northern European backgroundAmer J Hemat77384386

11 

Botas L, Freitas J, Barroca JP, Proença C, Pereira JMD, Martins e Silva JA1982Accion de la nifedipina en la respuesta al esfuerzo y en la oxigenacíon tisular periférica. Estudio comparativo com indivíduos sanos y pacientes de insuficiência coronáriaRev Espanhola Cardiologia35II75II81

12 

Levy Cruz F, Proença C, Freitas JP, Ramallho PS, Martins-Silva J1982Afinidade da hemoglobina para ooxigénio,2,3-difosfoglicerato e enzimas reguladoras da glicólise eritrocitária em estados diversosde retinopatia diabéticaRev Port Clin Terap78388

13 

Samaja M, Melotti D, Carenini A, Pozza G1982Glycosylated haemoglobins and the oxygen affinity of whole bloodDiabetologia23399402

14 

Martins-Silva J, Levy-Cruz F, Freitas JP, Souza-Ramalho P1984Blood filterability and oxygen hemoglobin affinity in diabetic patients with and without retinopathyActa Diabetol Lat21133139

15 

Castilho EM, Glass M, Manço JC2003The effects of 2,3 diphosphoglycerate, adenosine triphosphate, and glycosylated hemoglobin on the hemoglobin on the hemoglobin-oxigen affinity of diabetic patientsBrazilian J Medic Biol Res36731737

16 

Kynoch PAM, Lehmann H1977Rapid estimation (2 1/2hours) of glycosylated haemoglobin for routine purposeLancet31016

17 

Proença C, Martins JM, Martins e Silva J1983Microviscosidade da membrana eritrocitaria, hemoglobina glicosilada e glicoproteínas séricas em diabéticos não-insulino-depemdentesActa Medica Port4479482

18 

Hilério S, Saldanha C, Martins-Silva J2003An in vitro study of adrenaline effect on human erythrocyte properties in both gendersClin Hemorheol Microcirc288998

19 

Mesquita R, Pires I, Saldanha C, Martins-Silva J2001Effects of acetylcholine and spermineNONOate on erythrocyte hemorheologic and oxygen carrying propertiesClin Hemorheol Microcirc25153163

20 

Bogdanova A, Berenbrink M, Nikinmaa M2009Oxygen-dependent ion transport in erythrocytesActa Physiol (Oxf)195305319

21 

Tuvia S, Moses A, Gulayev N, Levin S, Korenstein R1999Beta-adrenergic agonists regulate cell membrane fluctiations of human erythrocytesJ Physiol516781792

22 

Campanella ME2005Assembley and regulation of a glycolytic enzyme on human erythrocyte membraneProc Nat Acad Sci10224022407

23 

Ince C2005Sidestream dark field imaging: An improved technique to observe sublingual microcirculationCrit CareSup 172

24 

De Backer D, Hollenberg S, Boerma C, Goedhart P, Büchele G, Ospina-Tascon G, Dobbe I, Ince C2007How to evaluate the microcirculation: Report of a round table conferenceCritical Care11R101

25 

Das L, Cohly H, Reno D, GoswamiIII, Das SKW1997Laser-Doppler evaluation of the human thympanic membrane by measuring blood flow volume and velocityIndian Journal of Otolaryngology & Head and Neck Surgery49132135

26 

Kunimoto Y, Hasegawa K, Arii S, Kataoka H, Yazama H, Kuya J, Kitano H2014Sequential multipoint motion of the tympanic membrane measured by laser Doppler vibrometry: Preliminary results for normal tympanic membraneOtol Neurotol35719724

27 

IJzerman RG, de Jongh RT, Beijk MA, van Weissenbruch MM, Delemarre-van de Waal HA, Serné EH, Stehouwer CD2003Individuals at increased coronary heart disease risk are characterized by impaired microvascular function in skinEur J Clin Invest33536542

28 

Bondesson SM, Edvinsson ML, Pettersson T, Edvinsson L2011Reduced peripheral vascular reactivity in refractory angina pectoris: Effect of enhanced external counterpulsationGeriatr Cardiol8215223

29 

Furchgott RF, Vanhoutte PM1989Endothelium-derived relaxing and contracting factorsFASEB J320072018

30 

Gow AJ, Luchsinger BP, Pawloski JR, Singel DJ, Stamler JS1999The oxyhemoglobin reaction of nitric oxideProc NatlAcad Sci U S A9690279032

31 

Z Huang, Ucer KB, Murphy T, Williams RT, King SB, Kim-Shapiro DB2002Kinetics of nitric oxide binding to R-state haemoglobinBiochem Biophys Res Commun292812818

32 

Chan NL, Kavanaugh JS, Rogers PH, Arnone A2004Crystallographic analysis of the interaction of nitric oxide with quaternary-T human hemoglobinBiochemistry43118132

33 

Azarov I, Huang KT, Basu S, Gladwin MT, Hog N, Kim-Shapiro DB2005Nitric oxide scavenging by red blood cells as afunction of hematocrit and oxygenationJ Biol Chem2801902419032

34 

Huang KT, Han TH, Hyduke DR, Vaughn MW, Herle HV, Hein TW, Zhang C, Kuo L, Liao JC2001Modulation of nitric oxidebioavailability by erythrocytesProc Natl Acad Sci U S A981177111776

35 

Ellsworth ML, Ellis CG, Goldman D, Stephenson AH, Dietrich HH, Sprague RS2009Erythrocytes: Oxygen Sensors and Modulators of Vascular Tone in Regions of Low PO2,Physiology24107116

36 

Lobisheva I, Biller P, Gallez B, Beauloye C, Blligand J2013Nitrosylated hemoglobinlevels in human venous erythrocytes correlate with vascular endothelial function measured by digital reactive hyperemiaPloS One810.1371/journal.pone.0076457

37 

Mesquita R, Saldanha C, Martins-Silva J2000Acetylcholine induces nitric oxide productionby erythrocytes in vitro Nitric Oxide4313314

38 

Kleinbongard P, Schulz R, Rassaf T, Lauer T, Dejam A, Jax T, Kumara I, Gharini P, Kabanova S, Ozüyaman B, Schnürch HG, Gödecke A, Weber AA, Robenek M, Robenek H, Bloch W, Rösen P, Kelm M2006Red blood cells express a functional endothelial nitric oxide synthaseBlood10729432951

39 

Ulker P, Yaras N, Yalcin O, Celik-Ozenci C, Johnson PC, Meiselman HJ, Baskurt OK2011Shear stress activation of nitric oxide synthase and increased nitric oxide levels in human red blood cellsNitric Oxide24173288

40 

Böhmer A, Beckmann B, Sandmann J, Tsikas D2012Doubts concerning functional nitric oxide synthase in human erythrocytesBlood11913221323

41 

Carvalho FA, Martins-Silva J, Saldanha C2004Amperometric measurements of nitric oxide in erythrocytesBiosens Bioelectron20505508

42 

Saldanha C, Lopes de Almeida JP, Silva-Herdade AS2014Application of a nitric oxide sensor in BiomedicineBiosensors4117

43 

Carvalho FA, Maria AV, Braz Nogueira JM, Guerra J, Martins-Silva J, Saldanha C2006The relation between the erythrocyte nitric oxide and hemorheological parametersClin Hemorheol Microcirc35341347

44 

Cook JA, Kim SY, Teague D, Krishna MC, Pacelli R, Mitchell JB1996Convenient colorimetric and fluorometric assays for S-nitrosothiolsAnal Biochem238150158

45 

Saldanha C, Teixeira P, Santos-Freitas T, Napoleão P2013Timolol Modulates Erythrocyte Nitric Oxide BioavailabilityJ Clin Exp Ophthalmol4285

46 

Woronzoff-Dashkoff KK2002The wright-giemsa stain. Secrets revealedClin Lab Med221523

47 

Oatley CW, Nixon WC, Pease RFW1965Scanning electron microscopyAdv Electronics Electron Phys21181247

48 

Markesbery WR, Butterfield DA1977Scanning electron microscopy studies of erythrocytes in Huntington‘s diseaseBiochem Biophys Res Comm78560564

49 

Arutjunov VD, Batsura JD, Gribova JA, Kruglikov GG1981Scanning electron-microscopic and light-optics investigations of erythrocytes in toxic anaemiaBritish J of Industrial Medicine387275

50 

Yasuda Y, Akiguchi I, Shio H, Kameyama M1984Scanning electron microscopy studies of erythrocytes in spinocerebellar degenerationJ Neurob Neuros Psych47269274

51 

“The Nobel Prize in Physics 1953”.Nobelprize.org.Nobel Media AB 2014, http://www.nobelprize.org/nobel_prizes/physics/laureates/1953/

{ label needed for ref[@id='r51a'] } 

Franke RP, Scharnweber T, Fuhrmann R, Wenzel F, Krüger A2014Effect of Radiographic Contrast Media on the Spectrin/Band3-Network of the Membrane Skeleton of ErythrocytesPLoS ONE9e8951210.1371/journal.pone.0089512

52 

“The Nobel Prize in Chemistry 2014”.Nobelprize.org.Nobel Media AB 2014 http://www.nobelprize.org/nobel_prizes/chemistry/laureates/2014/

53 

De Oliveira S, Vitorina de Almeida V, Calado A, Rosário HS, Saldanha C1818Integrin-associated protein (CD47) is a putative mediator for soluble fibrinogen interaction with human red blood cells membraneBiochem Biophys Acta481490

54 

Corsi D, Paiardini M, Crinelli R, Bucchini A, Magnani M1999Alteration of alpha-spectrin ubiquitination due to age-dependent changes in the erythrocyte membraneEur J Biochem261775783

55 

Venerando B, Fiorilli A, Croci G, Tringali C, Goi G, Mazzanti L, Curatola G, Segalini G, Massaccesi L, Lombardo A, Tettamanti G2002Acidic and neutral sialidase in the erythrocyte membrane of type 2 diabetic patientsBlood9910641070

56 

Saldanha C, Santos NC, Martins-Silva J2002Fluorescent probes DPH, TMA-DPH and C17-HC induce erythrocyte exovesiculationJ Membr Biol1907582

57 

Ellman GL, Courtney KD, Andres V, Featherstone RM1961A new and rapid colorimetry Determinationof acetylcholinesterase activityBiochem Pharmacol78895

58 

Saldanha C, Santos NC, Martins-Silva J2004A colorimetric process to visualize erythrocyte exovesicles aggregatesBiochem Mol Biol Educ32250253

59 

Kay MMB, Flowers N, Goodman J, Bosman G1989Alteration in membrane protein band 3 associated with acceleratederythrocyte agingProc Natl Acad Sci U S A8658345838

60 

Lang F, Gulbins E, Lerche H, Huber SM, Kempe DS, Föller M2008Eryptosis, a Window to systemic diseaseCell Physiol Biochem22373380

61 

de Jong K, Beleznay Z, Ott P1281Phospholipid asymmetry in red blood cells and spectrin free vesicles during peolonged storageBioch Biophys Acta101110

62 

Martins e Silva J, Carlota Proença M, Braz Nogueira J, Gorjão-Clara J, Nogueira da Costa J, Manso C1980Erythrocyte actylcholinesterase in essential hypertensionJ Neural Transm49127132

63 

Festoff BW, Fernandez HL1981Plasma and red blood cell acetylcholinesterase in amyotrophic lateral sclerosisMuscle Nerve44147

64 

Chow FL, Telen MJ, Rosse WF1985The acetylcholinesterase defect in paroxysmal nocturnal hemoglobinuria: Evidence that the enzyme is absent from the cell membraneBlood66940945

65 

Saldanha C. Acetylcholinesterase. Contribution for the kinetic study of the human erythrocyte enzyme, Ph.D. thesis (in Portuguese), 1985

66 

Carvalho FA, Almeida JP, Fernandes IO, Freitas-Santos T, Saldanha C2008Non-neuronal cholinergic system and signaltransduction pathways mediated by band 3 in red cellClin Hemorheol Microcirc40207227

67 

Carvalho FA, Almeida JP, Fernandes IO, Freitas-Santos T, Saldanha C2009Modulation of erythrocyte acetylcholinesterase activity and its association with G Protein Band 3 interactionsJ Membr Biol2288997

68 

Saldanha C, Freitas T, Almeida JP2012Fibrinogen effects on erythrocyte nitric oxide mobilixation in presence of acetylcholineLife Sciences9110171022

69 

Lopes de Almeida JP, Freitas-Santos T, Saldanha C2009Fibrinogen-dependent signalling microvascular erythrocyte function: Implications on nitric oxide effluxJ Memb Biol2314753

70 

Saldanha C, Freitas T, Almeida JP, Silva-Herdade A2014Signal transduction pathays in erythrocyte nitric oxide metabolismo under high fibrinogen levelsKorea-Australia Rheology J26217223

71 

Almeida JP, Freitas-Santos T, Saldanha C2011Evidence that the degree of band 3 phosphorylation Modulates human nitric oxide efflux-in vitro model of fibrinogenemiaClin Hemorheol Microcirc49407416

72 

Bordin L, Brunati AM, Donella-Deana A, Baggio B, Toninello A, Clari GBand 3 is an anchor protein and a target for SHP-2 tyrosine phosphatase in human erythrocytesBlood100202276282

73 

Saldanha C, Silva AS, Gonçalves S, Martins-Silva J2007Modulation of erythrocyte hemorheogical properties by band 3 phosphorylation and dephosphorylationClin Hemorheol Microcirc36183194

74 

Baskurt OK, Yyuklu M, Hardman MR, Meiselman HJ2009Photometric measurements of red blood cell aggregation: Light transmission versus light reflectanceJ Biom Opt14054044

75 

Zijlstra WG1958Syllectometry anew method for studying rouleaux formation of red blood cellsActa Physiol Pharmacol Nederl7153154

76 

Kiesewetter K, Radtke H, Schneider R, Mussler K, Scheffer A, Schmid-Schönbein H1982The mini erythrocyte aggregometer: A new apparatus for the rapid quantification of the extent of erythrocyte aggregationBiomed Tech28209213

77 

Baskurt OK, Meiselman HJ, Kayar E1998Measurements of red blood cell aggregation in a plate-plate shearing system by analysis of light transmissionClin Hemorheol Microc1930173314

78 

Popel AS, Johnson PC, Kaameneva MV1994Wild Capacity of red blood cell aggregation is higher in athletic mammalian species than in sedentarian speciesJ Appl Physiol7717901794

79 

Hardeman MR, Doble JGG, Ince C2001The laser-assisted optical rotational cell analyser (LORCA) as red blood cell aggregometerClin Hemorheol Microcirc25111

80 

Hardeman MR, Goedhart pT, Shin S 2007 Methods in hemorheologu Baskurt OK, Hardeman MR, Rampling mw, Meiselman HJ Handbook Hemorheology and Hemodynamics IOS Press Amsterdam, Berlin, Oxford, Tokyo, Washington DC 242 266 Eds

81 

Rampling MW, Meiselman HJ, Neu B, Baskurt OK2004Influence of cell-specific factors on red blood cell aggregationBiorheology4191112

82 

Baskurt OK, Bor-Kucukatay M, Yalcin O, Meiselman HJ, Armstrong JK2000Standard aggregation media to test the aggregability of rat blood cellsClin Hemorheol Microc22161166

83 

Baskurt OK, Boynard M, Cokelet GR, Connes P, Cooke BM, Forconi S, Liao F, Hardeman MR, Jung F, Meiselman HJ, Nash GB, Nemeth N, Neu B, Sandhagen B, Shin S, Thurston GB, Wautier JL2009New guidelines for hemorheological laboratory techniquesClin Hemorheol Microc227597

84 

Tubiana M, Amar MH, Burgers MV, van der Werf-Messing B, Hayat H1984Prognostic significance of erythrocyte sedimentation rate in clinical stages I-II of Hodgkin‘s diseaseAmerican Soc Clin Onc2194200

85 

Timmer JR, Ottervanger JP, Hoorntie JC, De Boer MJ, Survapranata H, van‘t Hof AW, Ziilstra F2005Myocardial Infarction study grouPrognostic value of erythrocyte sedimentation rate in ST segment elevation myocardial infarction: Interaction with hyperglycaemiaJ Intern Med257423429

86 

Wintrobe MM, Landsberg JW1975A standardized technique for blood sedimentation testAm j Med Sci189102

87 

Curves J, Kooren J, Laan M2010Evaluation of the Ves-Matic Cube 200 erythrocte sedimentation method: Comparasion with Westergren-based methodsAm J Clin Pathol134653660

88 

Bogdaycioglu N, Yilmaz FM, Sezer S, Oguz E2014Compararison of iSED and Ves-Matic Cube 200 sedimentation rate measurements with westergree methodJ Clin Lab Analys00116

89 

Cicco G, Vicenti P, Stingi GD, Tarallo , Pirrelli A1999Hemorheology in complicated hypertensionClin Hemorheol Microcirc21315319

90 

Martins e Silva J, Carlota Proença M, Braz Nogueira J, Gorjão-Clara J, Nogueira da Costa J, Manso C1980Erythrocyte actylcholinesterase in essential hypertensionJ Neural Transm49127132

91 

Konstantinova E, Ivanova L, Tolstaya T, Mironova E2006Rheological properties of blood and parameters of platelets aggregation in arterial hypertensionClin Hemorheol Microcirc35135138

92 

Uyuklu M, Meiselman HJ, Baskurt OK2009Effect of hemoglobin oxygenation level on red blood cell deformability and aggregation parametersClin Hemorheol Microcirc41179188

93 

Braz-Nogueira J, Freitas A, Moreira C, Saldanha C, Martins e Silva J, Souza-Ramalho P, Nogueira da Costa J1988Hipertensão em fase acelerada. Estudo cardiológico, oftalmológico e hemorreológicoRev Port Cardiol7305318

94 

Maharshak N, Arbel Y, Shapira I, Berliner S, Ben-Ami R, Yedgar S, Barshtein G, Dotan I2009Increased strength of erythrocyte aggregates in blood of patients with inflammatory bowel diseaseInflamm Bowel Dis15707713

95 

Poggi M, Palareti G, Biagi R, Legnani C, Parenti M, Babini AC, Coccheri C1994Prolonged very lowcalorie diet in highly obese subjects reduces plasma viscosity and red Cell aggregation but not fibrinogenInt J Obsesity18490496

96 

Jung F, Spitzer S, Kiesewetter H, Feldmann M, Kotischke G, Blum C, Wenzel E, Jutzler GA1986Comparative investigation of the microcirculation in patients with hypertension and healthy adultsKlin Wochenschr6495661

97 

Vaya A, Martinez M, Carmena R, Aznar J1993Red blood cell aggregation and primary hyperlipoproteinemiaThromb Res72119126

98 

Rainer C, Norris S, Haywood LJ, Meiselman HJ1989Blood rheology and RBc aggregation in patients with angina pectoris and a prior history of myocardial infarctionClin Hemorheol9923934

99 

Saldanha C, Lopes de Almeida JP2011Erythrocyte as a link between Basic and clinical researchClin Hemorhreol Microcirc49463472

100 

Baskurt OK, Gelmont D, Meiselman HJ1998Red blood cell deformability in sepsisAm J resp Crit Care Med157421427

101 

Baskurt OK, Temiz A, Meiselman HJ1997Red blood cell aggregation in experimental sepsisJ Lab Clin Med130183190

{ label needed for ref[@id='r101a'] } 

Jung F, Mrowietz C, Hiebl B, Franke RP, Pindur G, Sternitzky R2011Influence of rheological parameters on the velocity of erythrocytes passing nailfold capillaries in humansClin Hemorheol Microcirc48129139

102 

Copley AL1987On erythrocyte aggregation and disaggregationClin Hemorheol7314

103 

Cooke BM, Lim CT2007Mechanical and adhesive properties of healthy and diseased red blood cellsHandbook of hemorheology and hemodynamicsBaskurt OKIOS pressNetherlandseds

104 

Behnke O1970A comparative study of microtubules of disk-shaped blood cellsJ Ultrastruct Res316175

105 

Cluitmans JCA, Chokkalingam V, Janssen AM, Brock R, Huck WTS, Bosman GJCGM2014Alterations in Red Blood Cell Deformability during Storage: A Microfluidic ApproachBioMed Research International2014Article ID 764268, 9 pages

106 

Reid HL, Barnes AJ, Look PJ, Dormandy JA1976A simple method for measuring erythrocyte deformabilityJ Clin Path29855

107 

Sakuta S1981Blood filterability in cerebrovascular disorders, with special reference to erythrocyte deformability and AtP contentStroken12824828

108 

Teitel P1977Basic Principles of the filtrability test (FT) and analysis of erythrocyte flow behaviourBlood Cells35570

109 

Seiffge D1984Correlation of different methods to determine red blood cell deformabilityClin Hemorheol4263273

110 

Schmid-Schönbein H, Wells R, Schildkraut R1969Microscopy and viscometry of blood flowing under uniform shear rateJ Appl Physiol26674

111 

Dobbe JGG, Streekstra MR, Hardeman MR, Ince C, Grimbergen CA2002Measurement of the distribution of red cell deformability using an automated rheoschopeCytometry50373384

112 

Antonova N2009Quantification and technique of measurements of RBC aggregation and deformabilityBoletim da SPHM24517

113 

Groner W, Mohandas N, Bessis M1980New optical technique for measuring erythrocyte deformability with the ektacytometerClin Chem2614351442

114 

Mohandas N, Clark NMR, Jacobs MS, Shohet SB1979Quantitative analysis of factors regulating erythrocyte deformabilityBlood54Suppl 130

115 

Prado G, Farutin A, Misbah C, Bureau L. Viscoelastic transient of confined red blood cells. arXiv: 1409.5049v1 [cond-mat.soft]

116 

Uyuklu M, Meiselman HJ, Baskurt OK2009Role of hemoglobin oxygenation in the modulation of red blood cell mechanical properties by nitric oxideNitric Oxide212026

117 

de Oliveira S, Silva-Herdade AS, Saldanha C2008Modulation of erythrocyte deformability by PKC activityClin Hemorheol Microcirc39363373

118 

Mannot S, Takakuwa Y, Mohandas N2005Modulation of erythrocyte membrane mechanical function by protein 4.1 phosphorylationJ Biol Chem28075817587

119 

Saldanha C, Almeida JP, Freitas T, Oliveira S, Silva-Herdade AS2010Erythrocyte deformability responses to shear stress under external and internal stimuli influencesSeries in Biomech255460

120 

Silva-Herdade AS, Freitas T, Almeida JP, Saldanha C2014Erythrocyte deformability and nitric oxide mobilizationunder pannexin-1 and PKC dependenceClin Hemorheol Microc(Epub ahead of print]

121 

Keymel S, Heiss C, Kleinbongard P, Kelm M, Lauer T2011Impaired red blood cell deformability in patients with coronary artery disease and diabetes mellitusHorm Metab Res43760765

122 

Shin S, Ku Y2005Hemorheology and clinical application: Association of impairment of red blood cell deformability with diabetic nephropathyKorea-Australia Rheology J17117123

123 

Coulter WH1956High speed automatic blood cell counter and size analyzerProc Natl Electron Conf10341042

124 

Lippi G, Salvagno GL, Danese E, Tarperi C, Guidi GC, Schena F2014Variation of red blood cell distribution widthand mean platelet volume after moderate endurance exerciseAdvances HematolArticle ID 192173, 4 pageshttp://dx.doi.org/10.1155/2014/192173

125 

Uyarel H, Isik T, Ayhan E, Ergelen M2011Red cell distribution width (RDW): A novel risk factor for cardiovascular diseaseInt J Card504047

126 

Tomaiuolo G, Guido S2011Start-up shape dynamics of red blood cells in microcapillary flowMicrovasc Res823541

127 

Tomaiuolo G, Rossi D, Caserta S, Cesarelli M, Guido S2012Comparation of two flow-based imaging methods to measure individual red blood Cell área and volumeCitometry81A10401047