Article type: Research Article
Authors: Wilke, Jan* | Vogt, Lutz | Pfarr, Tatjana | Banzer, Winfried
Affiliations: Department of Sports Medicine, Goethe University Frankfurt am Main, Frankfurt am Main, Germany
Correspondence: [*] Corresponding author: Jan Wilke, Department of Sports Medicine, Goethe University Frankfurt am Main, Ginnheimer Landstraße 39, D-60487 Frankfurt am Main, Germany. Tel.: +49 69 798 245 88; Fax: +49 69 798 245 82; E-mail: [email protected].
Abstract: BACKGROUND:Most methods assessing soft tissue stiffness are associated with high costs and considerable technical effort. A previous study demonstrated that the tissue compliance meter represents an affordable alternative with straightforward handling. However, little data exists describing its measurement properties. OBJECTIVE:The study aimed to elucidate the reliability and validity of a semi-electronic tissue compliance meter (STCM). METHODS:Tissue stiffness at the lateral calf was examined in 34 healthy participants (24 ± 5 yrs, 21 men). Measurements on two consecutive days were used to estimate test-retest reliability, day-to-day reliability and inter-rater reliability. Validity was examined twofold. Firstly, STCM data were correlated with those of another method, the myotonometry. Secondly, STCM measurements at different levels of isometric contraction (M.gastrocnemius) were examined for systematic differences. RESULTS:Intraclass correlations (ICC) revealed high test-retest reliability (ICC 0.84, 95% CI: 0.70 to 0.92; p< 0.05) and day-to-day reliability (0.84, 0.71 to 0.92; p< 0.05). Inter-rater reliability was sufficient to good (0.75, 0.56 to 0.87; p< 0.05). Cross-validation with myotonometry revealed a strong correlation of both methods (r= 0.86; p< 0.05). Stiffness assessments during isometric contraction demonstrated that the STCM can discriminate different force levels (χ2= 57.1, p< 0.05). CONCLUSIONS:The STCM provides valid and reliable measurements of tissue stiffness. Further study is needed to provide norm values for different body regions and to reveal alterations in patients with musculoskeletal disorders.
Keywords: Soft tissue, tissue hardness, elasticity, muscle tone
DOI: 10.3233/BMR-170871
Journal: Journal of Back and Musculoskeletal Rehabilitation, vol. 31, no. 5, pp. 991-997, 2018
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]