Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Simmonds, Michael J.a; * | Meiselman, Herbert J.b
Affiliations: [a] Menzies Health Institute Queensland, Griffith University, QLD, Australia | [b] Department of Physiology and Biophysics, Keck School of Medicine, University of Southern California, Los Angeles, CA, USA
Correspondence: [*] Address for correspondence: Michael J. Simmonds, Menzies Health Institute Queensland, Griffith University, QLD, Australia. Tel.: +61 (7) 5552 8529; Fax: +61 (7) 5552 8674; E-mail: [email protected].
Abstract: Background:Current generation mechanical circulatory assist devices are designed to minimize high shears to blood for prolonged durations to avoid hemolysis. However, red blood cells (RBC) demonstrate impaired capacity to deform when exposed to shear stress (SS) well below the “hemolytic threshold”. Objective:We endeavored to identify how changes in the magnitude and duration of SS exposure alter RBC deformability and subsequently develop a model to predict erythrocyte subhemolytic damage. Methods:RBC suspensions were exposed to discrete magnitudes of SS (1–64 Pa) for specific durations (1–64 s), immediately prior to RBC deformability being measured. Analyses included exploring the maximal RBC deformation (EImax) and SS required for half EImax (SS1/2). A surface-mesh was interpolated onto the raw data to predict impaired RBC deformability. Results:When SS was applied at <16Pa, limited changes were observed. When RBC were exposed to 32 Pa, mild impairments in EImax and SS1/2 occurred, although 64 Pa caused a dramatic impairment of RBC deformability. A clear relation between SS duration and magnitude was determined, which could predict impaired RBC deformability. Conclusion:The present results provide a model that may be used to predict whether RBC deformability is decreased following exposure to a given level and duration of SS, and may guide design of future generations of mechanical circulatory assist devices.
Keywords: Blood damage, haemocompatibility, haemorheology, mechanical damage, sublethal
DOI: 10.3233/BIR-16120
Journal: Biorheology, vol. 53, no. 5-6, pp. 237-249, 2016
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]