Article type: Research Article
Authors: Singh, Om Prakasha; * | Ahmed, Ismail Binb | Malarvili, M.B.a
Affiliations: [a] Bio-signal Processing Research Group, Faculty of Biosciences and Medical Engineering, Universiti Teknologi Malaysia, Skudai 81310, Malaysia | [b] Pengarah Pusat Kesihatan Mahasiswa, Universiti Teknologi Malaysia, Skudai 81310, Malaysia
Correspondence: [*] Corresponding author: Om Prakash Singh, Bio-signal Processing Research Group (BSPRG), Faculty of Biosciences and Medical Engineering, Universiti Teknologi Malaysia (UTM), Skudai 81310, Johor Bahru, Johor, Malaysia. Tel.: +60 7 5535982; E-mail: [email protected].
Abstract: BACKGROUND: Assessment of asthma outside of the hospital using a patient independent device is highly in demand due to the limitation of existing devices, which are manual and unreliable if patients are not cooperative. OBJECTIVE: The study aims to verify the use of newly developed human respiration, carbon dioxide (CO2) measurement device for the management of asthma outside of the hospital. METHOD: The data were collected from 60 subjects aged between 18–35 years via convenience sampling method reported in UTM Health Center using the device. Furthermore, the data were normalized and analyzed using descriptive statistics, t-test, and area (Az) under receiver operating characteristic curve (ROC). RESULT: Findings revealed that the normalized mean values of end-tidal carbon dioxide (EtCO2), Hjorth Activity (HA), and respiratory rate (RR) were lower in asthmatic compared with healthy subjects with minimum deviation from the mean. In addition, each parameter was found to significantly differ statistically for asthma and non-asthma with p< 0.05. Furthermore, the Az shows the strong association for the screening of asthma and non-asthma with an average of 0.71 (95% CI: 0.57–0.83), 0.77 (95% CI: 0.64–0.90), and 0.83 (95% CI: 0.73–0.94) for RR, EtCO2, and HA, respectively. CONCLUSIONS: This study demonstrates that the newly developed handheld human respiration CO2 measurement device may possibly be used as an effort-independent asthma management method outside of the hospital.
Keywords: Asthma, capnography, early screening, CO2 measurement
DOI: 10.3233/THC-181288
Journal: Technology and Health Care, vol. 26, no. 5, pp. 785-794, 2018
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]