Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Laib, Andresa; * | Häuselmann, Hans Jörgb | Rüegsegger, Petera
Affiliations: [a] Institute for Biomedical Engineering, University of Zürich and Swiss Federal Institute of Technology (ETH), Moussonstrasse 18, CH-8044 Zürich, Switzerland | [b] Departement of Rheumatology, University Hospital, Gloriastrasse 25, CH-8091 Zürich, Switzerland
Correspondence: [*] Address for correspondence: Andres Laib, Institute for Biomedical Engineering, Moussonstrasse 18, CH-8044 Zürich, Switzerland. Tel.: +41 1 632 45 92; Fax: +41 1 632 12 14; E-mail: [email protected].
Abstract: In vivo examinations of bone microarchitecture have become available recently through high resolution computed tomography (3D-QCT) and magnetic resonance imaging. The spatial resolution of the resulting images, however, is not sufficient to depict individual trabeculae in their true shape. Nevertheless, structural indices such as relative bone volume, trabecular number, mean thickness and mean separation can be extracted with the help of a ridge detection algorithm. Precision of the procedure is of the order of 1%, accuracy is ascertained using a micro-CT based calibration. In this work we report first results of time serial examinations. Eighteen healthy postmenopausal women (no HRT) were measured at months 0, 6, and 12, and the temporal changes were analyzed. Examination site was the distal radius. The above mentioned structural indices, the average densities and the thickness of the cortical shell were determined. Of the 18 women 6 showed no significant bone loss of any kind, 5 lost primarily cancellous bone, 4 lost primarily cortical bone, and 3 had a substantial loss of cortical as well as cancellous bone. We conclude that even in a homogenous group such as postmenopausal women, there are considerable differences in the reason why bone is weakend and that high resolution 3D-QCT allows to differentiate between various types of bone loss.
Keywords: Computed tomography (CT), bone microarchitecture, trabecular bone structure, quantitative bone morphology
DOI: 10.3233/THC-1998-65-606
Journal: Technology and Health Care, vol. 6, no. 5-6, pp. 329-337, 1998
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]