Article type: Research Article
Authors: Wiebking, Ulricha; * | Birkenhauer, Björnb | Krettek, Christiana | Gösling, Thomasa
Affiliations: [a] Department of Trauma Surgery, Hannover Medical School, Hannover, Germany | [b] Department of Orthopaedic Surgery, Diakoniekrankenhaus Bad Kreuznach, Bad Kreuznach, Germany
Correspondence: [*] Address for correspondence: Dr. Ulrich Wiebking, Department of Traumatology, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany. Tel.: +49 511/532 3802; Fax: +49 511/532 8823; E-mail: [email protected].
Abstract: Background:Use of the proximal part of the femur in total hip arthroplasty enables preservation of the distal femur for later revisions. To use this advantage, different types of short-stem prosthesis have been developed in recent years. Although cementless hip arthroplasty is not common in the treatment of canine osteoarthritis, the use of cementless short-stems might be an alternative therapy. The new cementless short-stem prosthesis called Spiron® is self-tapping, and is constructed with a conical shape with threads. We measured the relative motion in the bone/prosthesis interface with specified loads in the femora of dogs to investigate two aspects: the primary stability of two systems of uncemented prosthesis with different principles of anchoring, and the theoretical use of the Spiron® in dog bone. We measured the cyclic behaviour (i.e., reversible, elastic), subsidence (i.e., irreversible, plastic, migration) and maximal applied load. Methods:Twenty-four pairs of fresh femur bones from adult German shepherd dogs were used. After measuring the total bone mineral density (TBMD), 16 bones were used in each of the short-stem prosthesis group (group A), the Zweymuller prosthesis group (group B), and the no-prosthesis control group (group C). Micromotion between bone and prostheses was measured for 16,200 N axial load steps, beginning with 200 N and increasing to 3000 N (1600 cycles/femur). Simple analysis of variance and non-parametric tests were used to compare the groups. Results:The Spiron prosthesis had significantly less motion in the bone/prosthesis interface compared with the Zweymuller prosthesis. Conclusions:The new principle of anchoring of the Spiron short-stem prosthesis may provide higher primary stability compared with conventional techniques. The findings of this study support the assumption that the use of the Spiron prosthesis to treat osteoarthritis in the dog is feasible.
Keywords: Short-stem prosthesis, proximal anchorage, primary stability, in vitro study, canine
DOI: 10.3233/THC-2011-0634
Journal: Technology and Health Care, vol. 19, no. 4, pp. 271-282, 2011
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]