Article type: Research Article
Authors: Kouloulias, V.E.a; c; * | Poortmans, P.b | Antypas, C.c | Kappas, C.d | Sandilos, P.c
Affiliations: [a] European Organization for Research and Treatment of Cancer (EORTC), Data Center, Av. Mounier 83, B-1200, Brussels, Belgium | [b] Dr. Bernard Verbeeten Instituut, Dept. of Radiotherapy, Tilburg, The Netherlands | [c] Department of Radiotherapy, Areteion University Hospital, Medical School, Vas. Sofias St. 76, 115 28, Athens, Greece | [d] Department of Medical Physics, University of Thessalia, Larisa, Greece
Correspondence: [*] Address for correspondence: Dr. Vassilis Kouloulias (MS, MD, PhD), EORTC, Radiotherapy group, Quality Assurance, Data Center, Av. Mounier 83, B-1200, Brussels, Belgium. E-mail: [email protected].
Abstract: Flatness and symmetry are main parameters determining the quality of a photon beam produced by linear accelerators. The quality assurance in routine clinical practice of radiotherapy and consequently the treatment-outcome depend definitely on the physical parameters of treatment-delivery. Several recommendations from national and international associations are reported. By reviewing the current literature and mainly according to the World Health Organization (WHO) report of quality assurance in radiotherapy, we may suggest that for flatness and symmetry, the optimal level of deviation should be within ± 3%. Flatness and symmetry should be checked monthly or once a year in accordance to the guidelines of national societies.
Keywords: photon beam, quality assurance, symmetry and flatness, recommendations, review
DOI: 10.3233/THC-2003-11408
Journal: Technology and Health Care, vol. 11, no. 4, pp. 283-288, 2003
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]