Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Donovan Jr., Francis M.a | Taylor, Bruce C.b
Affiliations: [a] Mechanical Engineering Department, University of South Alabama, EGCB 212, Mobile, AL 36688, USA. Tel.: +1 334 460 6168; Fax: +1 334 460 6549; E-mail: [email protected] | [b] Biomedical Engineering Department, The University of Akron, Akron, OH 44325, USA
Abstract: Cardiac output is measured by the thermal dilution method which uses a quadruple lumen catheter, with a thermistor on the tip, through the right atrium, right ventricle and into the pulmonary artery. Cold saline is injected into the right atrium and the resulting pulmonary artery temperature profile is integrated. The same procedure performed with three thermistors and three pressure sensors located on the catheter to measure temperature and pressure in the atrium, ventricle and artery respectively will produce a set of temperature and pressure curves with shapes determined by injectate temperature, injectate volume, heart rate, systolic time interval, body temperature, cardiac output, volumes, flow rates and valve openings. A digital computer program has been developed to optimize the fit of a lumped parameter model to the thermodilution curves in order to determine heart rate, systolic time as a fraction of cardiac cycle, right atrial systolic and diastolic volumes, ventricular systolic and diastolic volumes, cardiac output, inflow valve forward and reverse flow rates and effective diameters, outflow valve forward and reverse flow rates and effective diameters, ventricular power and efficiency. The program has been tested over a range of operating conditions including noise in the temperature and pressure signals, randomly varying heart rate and cardiac cycle. All of the data for the tests were produced by a digital computer simulation of a pulsatile artificial heart. The results of these tests indicate that the enhanced thermal dilution analysis method is feasible.
DOI: 10.3233/THC-2001-9501
Journal: Technology and Health Care, vol. 9, no. 5, pp. 377-386, 2001
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]