Article type: Research Article
Authors: Lau, Andy | Ahmad, Salahuddin | Chen, Yong*
Affiliations: Department of Radiation Oncology, Peggy and Charles Stephenson Cancer Center, The University of Oklahoma Health Sciences Center, Oklahoma City, OK, USA
Correspondence: [*] Corresponding author: Yong Chen, Department of Radiation Oncology, Peggy and Charles Stephenson Cancer Center, The University of Oklahoma Health Sciences Center, 800 N.E. 10th St., OKCC L100, Oklahoma City, OK 73104, USA. Tel.: +1 405 271 3220; Fax: +1 405 271 8297; E-mail: [email protected].
Abstract: In vivo range verification methods will reveal information about the penetration depth into a patient for an incident proton beam. The prompt gamma (PG) method is a promising in vivo technique that has been shown to yield this range information by measuring the escaping MeV photons given a suitable detector system. The majority of current simulations investigating PG detectors utilize common scintillating materials ideal for photons within a low neutron background radiation field using complex geometries or novel designs. In this work we simulate a minimal detector system using a material ideal for MeV photon detection in the presence of a significant neutron field based on the Cherenkov phenomenon. The response of this selected material was quantified for the escaping particles commonly found in proton therapy applications and the feasibility of using the PG technique for this detector material was studied. Our simulations found that the majority of the range information can be determined by detecting photons emitted with a timing window less than ∼50 ns after the interaction of the proton beam with the water phantom and with an energy threshold focusing on the energy range of the de-excitation of 16O photons (∼6 MeV). The Cherenkov material investigated is able to collect these photons and estimate the range with timescales on the order of tens of nanoseconds as well as greatly suppress the signal due to neutron.
Keywords: Proton therapy, GEANT4, Monte Carlo, Cherenkov, prompt gamma
DOI: 10.3233/XST-160575
Journal: Journal of X-Ray Science and Technology, vol. 24, no. 4, pp. 565-582, 2016
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]