Article type: Research Article
Authors: Zhou, Xuana; b | Liu, Yuedonga; b | Wei, Cunfenga; b; c | Xu, Qionga; c; *
Affiliations: [a] Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Science, Beijing, China | [b] School of Nuclear Science and Technology, University of Chinese Academy of Sciences, Beijing, China | [c] Jinan Laboratory of Applied Nuclear Science, Jinan, China
Correspondence: [*] Corresponding author: Qiong Xu, Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Science, Beijing 100049, China. E-mail: [email protected].
Abstract: BACKGROUND:Geometry calibration for robotic CT system is necessary for obtaining acceptable images under the asynchrony of two manipulators. OBJECTIVE:We aim to evaluate the impact of different types of asynchrony on images and propose a reference-free calibration method based on a simplified geometry model. METHODS:We evaluate the impact of different types of asynchrony on images and propose a novel calibration method focused on asynchronous rotation of robotic CT. The proposed method is initialized with reconstructions under default uncalibrated geometry and uses grid sampling of estimated geometry to determine the direction of optimization. Difference between the re-projections of sampling points and the original projection is used to guide the optimization direction. Images and estimated geometry are optimized alternatively in an iteration, and it stops when the difference of residual projections is close enough, or when the maximum iteration number is reached. RESULTS:In our simulation experiments, proposed method shows better performance, with the PSNR increasing by 2%, and the SSIM increasing by 13.6% after calibration. The experiments reveal fewer artifacts and higher image quality. CONCLUSION:We find that asynchronous rotation has a more significant impact on reconstruction, and the proposed method offers a feasible solution for correcting asynchronous rotation.
Keywords: Robotic CT, reference-free, geometry calibration
DOI: 10.3233/XST-240023
Journal: Journal of X-Ray Science and Technology, vol. 32, no. 5, pp. 1239-1252, 2024
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]