Article type: Research Article
Authors: Zhang, Tianchenga | Almajidi, Yasir Q.b; * | Awad, Sameer A.c | Alhachami, Firas Rahid | Gatea, Maher Abdulfadhile | Kadhum, Wesam R.f
Affiliations: [a] Department of Radiology, The First People’s Hospital of Lianyungang, Lianyungang, China | [b] Baghdad College of Medical Sciences, Baghdad, Iraq | [c] Department of Medical Laboratory Techniques, Al-Maarif University College, Al-Anbar-Ramadi, Iraq | [d] Radiology Department, College of Health and Medical Technology, Al-Ayen University, Thi-Qar, Iraq | [e] Technical Engineering Department College of Technical Engineering, The Islamic University, Najaf, Iraq | [f] Department of Pharmacy, Kut University College, Kut, Wasit, Iraq
Correspondence: [*] Corresponding author: Yasir Q. Almajidi, Baghdad College of Medical Sciences, Baghdad, Iraq. E-mail: [email protected].
Abstract: BACKGROUND:Several physical factors such as photon beam energy, electron beam energy, and dose rate may affect the dosimetric properties of polymer gel dosimeters. The photon beam energy and dose rate dependence of PASSAG gel dosimeter were previously evaluated. OBJECTIVE:This study aims to assess the dosimetric properties of the optimized PASSAG gel samples in various electron beam energies. METHODS:The optimized PASSAG gel samples are first fabricated and irradiated to various electron energies (5, 7, 10 and 12 MeV). Then, the response (R2) and sensitivity of gel samples are analyzed by magnetic resonance imaging technique at a dose range of 0 to 10 Gy, scanning room temperature range of 15 to 22 °C, and post-irradiation time range of 1 to 30 days. RESULTS:The R2-dose response and sensitivity of gel samples do not change under the evaluated electron beam energies (the differences are less than 5%). Furthermore, a dose resolution range of 11 to 38 cGy is obtained for the gel samples irradiated to different electron beam energies. Moreover, the findings show that the R2-dose response and sensitivity dependence of gel samples on electron beam energy varies over different scanning room temperatures and post-irradiation times. CONCLUSION:The dosimetric assessment of the optimized PASSAG gel samples provides the promising data for this dosimeter during electron beam radiotherapy.
Keywords: Electron beam radiotherapy, PASSAG, MRI, Polymer gel dosimetry
DOI: 10.3233/XST-230073
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 4, pp. 825-836, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]