Authors: Ghani, Muhammad Usman | Makeev, Andrey | Manus, Joseph A. | Glick, Stephen J. | Ghammraoui, Bahaa; *
Affiliations: Division of Imaging, Diagnostics and Software Reliability, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, United States Food and Drug Administration, Silver Spring, MD, USA
Correspondence: [*] Corresponding author: Bahaa Ghammraoui, Ph.D. Senior Staff Fellow, Division of Imaging, Diagnostics and Software Reliability, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, United States Food and Drug Administration, Silver Spring, MD 20993, USA. Tel.: +1 301 796 3002; E-mail: [email protected].
Abstract: BACKGROUND:Geometric calibration is essential in developing a reliable computed tomography (CT) system. It involves estimating the geometry under which the angular projections are acquired. Geometric calibration of cone beam CTs employing small area detectors, such as currently available photon counting detectors (PCDs), is challenging when using traditional-based methods due to detectors’ limited areas. OBJECTIVE:This study presented an empirical method for the geometric calibration of small area PCD-based cone beam CT systems. METHODS:Unlike the traditional methods, we developed an iterative optimization procedure to determine geometric parameters using the reconstructed images of small metal ball bearings (BBs) embedded in a custom-built phantom. An objective function incorporating the sphericities and symmetries of the embedded BBs was defined to assess performance of the reconstruction algorithm with the given initial estimated set of geometric parameters. The optimal parameter values were those which minimized the objective function. The TIGRE toolbox was employed for fast tomographic reconstruction. To evaluate the proposed method, computer simulations were carried out using various numbers of spheres placed in various locations. Furthermore, efficacy of the method was experimentally assessed using a custom-made benchtop PCD-based cone beam CT. RESULTS:Computer simulations validated the accuracy and reproducibility of the proposed method. The precise estimation of the geometric parameters of the benchtop revealed high-quality imaging in CT reconstruction of a breast phantom. Within the phantom, the cylindrical holes, fibers, and speck groups were imaged in high fidelity. The CNR analysis further revealed the quantitative improvements of the reconstruction performed with the estimated parameters using the proposed method. CONCLUSION:Apart from the computational cost, we concluded that the method was easy to implement and robust.
Keywords: Geometric calibration, sphericity, symmetry, photon counting detectors, cone beam CTs
DOI: 10.3233/XST-230007
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 5, pp. 865-877, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]