Article type: Research Article
Authors: Sun, Yanmina | Han, Yua | Tan, Siyua | Xi, Xiaoqia | Li, Leia | Yan, Bina; * | Zhang, Yuanb
Affiliations: [a] Henan Key Laboratory of Imaging and Intelligent Processing, PLA Strategy Support Force Information Engineering University, Zhengzhou, Henan, China | [b] College of Information Science and Engineering, Henan University of Technology, Zhengzhou, Henan, China
Correspondence: [*] Corresponding author: Bin Yan, Henan Key Laboratory of Imaging and Intelligent Processing, PLA Strategy Support Force Information Engineering University, Zhengzhou, Henan, China. E-mail: [email protected].
Abstract: BACKGROUND:X-ray cone-beam computed laminography (CL) is widely used for large flat objects that computed tomography (CT) cannot investigate. The rotation angle of axis tilt makes geometric correction of CL system more complicated and has more uncertain factors. Therefore, it is necessary to evaluate sensitivity of the geometric parameters of CL system in advance. OBJECTIVE:This study aims to objectively and comprehensively evaluate sensitivity of CL geometric parameters based on the projection trajectory. METHODS:This study proposes the Minimum Deviation Unit (MDU) to evaluate sensitivity of CL geometric parameters. First, the projection trajectory formulas are derived according to the spatial relationship of CL system geometric parameters. Next, the MDU of the geometric parameters is obtained based on the projection trajectories and used as the evaluation index to measure the sensitivity of parameters. Then, the influence of the rotation angle of the axis tilt and magnification on the MDU of the parameters is analyzed. RESULTS:At low magnification, three susceptible parameters (η, u0, v0) with MDU less than 1 (° or mm) must be calibrated accurately to avoid geometric artifacts. The sensitivity of CL parameters increases as the magnification increases, and all parameters become highly sensitive when the magnification power is greater than 10. CONCLUSION:The results of this study have important guiding significance for the subsequent further parameter calibration.
Keywords: Cone-beam, computed laminography, X-ray laminography, geometric parameters, sensitivity evaluation, parameters sensitivity
DOI: 10.3233/XST-221338
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 2, pp. 423-434, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]