Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Sun, Jihanga | Li, Haoyana | Yang, Lixina | Zhou, Zuofub | Li, Michellec | Peng, Yuna; *
Affiliations: [a] Imaging Center, Beijing Children’s Hospital, Capital Medical University, National Center for Children’s Health, Beijing, China | [b] Department of Radiology, Fujian Provincial Maternity and Children’s Hospital, Affiliated Hospital of Fujian Medical University, Fujian, China | [c] Department of Human Biology, Stanford University, Stanford, CA, USA
Correspondence: [*] Corresponding author: Yun Peng, Imaging Center, Beijing Children’s Hospital, Capital Medical University and National Center for Children’s Health, 56 Nanlishi Road, Xicheng District, Beijing 100045, China. Tel.: +86 010 59617038; E-mail: [email protected].
Abstract: BACKGROUND:Low-tube voltage scanning improves CT attenuation value of contrast medium (CM). Thus, we hypothesized that 70 kVp in pediatric abdominal CT angiography (CTA) could be used to reduce both radiation and CM dose and improve patient comfort at the same time. OBJECTIVE:To evaluate the feasibility of using 70 kVp in pediatric abdominal CTA to reduce radiation dose and CM dose and improve patient care for children. MATERIALS AND METHODS:Forty-six children needing abdominal CTA were enrolled in the study group using low-dose scanning protocol with 70 kVp and 0.7–1.1 ml/kg contrast dose, and reconstructed with 50%ASIR-V. They were compared with other 46 children in control group with matching body weight and underwent conventional CT scans with 100 kVp, 1.2–1.8 ml/kg contrast dose and reconstructed using 50%ASIR. Image quality of large vessels was evaluated using a 5-point scale. CT value and standard deviation of descending aorta (Ao) was measured, and signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Radiation dose, contrast dose, the maximum injection pressure between the two groups were also compared. RESULTS:Score for displaying large vessels by 70 kVp images was 3.91±0.28, lower than that (4.17±0.38) of the control group (p < 0.05), but fully met the diagnostic requirements. CT value of Ao was 390.87±86.79HU in study group, which is higher than 343.93±49.94HU in control group, while there was no difference in SNR and CNR between two groups; the radiation dose, contrast dosage and injection pressure of the study group were 1.23±0.39mGy, 12.67±7.27 ml and 43.83±17.16psi, respectively, which are significantly lower than the 1.95±0.37mGy, 22.67±7.39 ml, and 77.59±19.68psi of control group. CONCLUSION:Use of 70 kVp in pediatric abdominal CTA provides diagnostic quality images while significantly reduce radiation and contrast dose, as well as injection pressure to improve patient comfort for children.
Keywords: Computed tomography angiography (CTA), X-ray imaging, low voltage of X-ray tube, contrast medium, children
DOI: 10.3233/XST-210896
Journal: Journal of X-Ray Science and Technology, vol. 29, no. 5, pp. 813-821, 2021
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]