Article type: Research Article
Authors: Chang, Chia-Hao; * | Ni, Yu-Ching | Tseng, Sheng-Pin
Affiliations: Health Physics Division, Institute of Nuclear Energy Research, Taoyuan City, Taiwan, R.O.C
Correspondence: [*] Corresponding author: Chia-Hao Chang, Health Physics Division, Institute of Nuclear Energy Research, No. 1000, Wenhua Rd. Jiaan Village, Longtan District, Taoyuan City 32546, Taiwan, R.O.C. Tel.: +886 3471 1400 #7763; Fax: +886 3471 1171; E-mail: [email protected].
Abstract: The study aims to develop a rational polynomial approximation method for improving the accuracy of the effective atomic number calculation with a dual-energy X-ray imaging system. This method is based on a multi-materials calibration model with iterative optimization, which can improve the calculation accuracy of the effective atomic number by adding a rational term without increasing the computation time. The performance of the proposed rational polynomial approximation method is demonstrated and validated by both simulated and experimental studies. The twelve reference materials are used to establish the effective atomic number calibration model, and the value of the effective atomic numbers are between 5.444 and 22. For the accuracy of the effective atomic number calculation, the relative differences between calculated and experimental values are less than 8.5%for all sample cases in this study. The average calculation accuracy of the method proposed in this study can be improved by about 40%compared with the conventional polynomial approximation method. Additionally, experimental quality assurance phantom imaging result indicates that the proposed method is compliant with the international baggage inspection standards for detecting the explosives. Moreover, the experimental imaging results reveal that the difference of color between explosives and the surrounding materials is in significant contrast for the dual-energy image with the proposed method.
Keywords: Effective atomic numbers, rational polynomial approximation, dual-energy X-ray, baggage inspection
DOI: 10.3233/XST-200790
Journal: Journal of X-Ray Science and Technology, vol. 29, no. 2, pp. 317-330, 2021
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]