Article type: Research Article
Authors: Zhang, Mutiana; * | Fan, Qiyongb | Lei, Yuc | Thapa, Bishnua | Padula, Gilberta
Affiliations: [a] Radiation Therapy, Summa Health Cancer Institute, Akron, Ohio, USA | [b] Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA | [c] Department of Radiation Oncology, University of Nebraska Medical Center, Omaha, Nebraska, USA
Correspondence: [*] Corresponding author: Mutian Zhang, PhD, Radiation Therapy, Summa Health Cancer Institute, 161 N Forge Street, Akron, Ohio, OH 44304, USA. Tel.: +1 330 375 7868; E-mail: [email protected].
Abstract: BACKGROUND:Versa HD linear accelerators (linacs) are used for stereotactic radiosurgery treatment. However, the mechanical accuracy of such systems remains a concern. OBJECTIVE:The purpose of this study was to evaluate the accuracy of an Elekta Versa HD linac. METHODS:We performed measurements with a ball bearing phantom to calculate the rotational isocenter radii of the linac’s gantry, collimator, and table, and determine the relative locations of those isocenters. We evaluated the accuracy of the cone-beam computed tomography (CBCT) guidance with a film-embedding head phantom and circular cone-collimated radiation beams. We also performed dosimetric simulations to study the effects of the linac mechanical uncertainties on non-coplanar cone arc delivery. RESULTS:The mechanical uncertainty of the linac gantry rotation was 0.78 mm in radius, whereas that of the collimator and the table was <0.1 mm and 0.33 mm, respectively. The axes of rotation of the collimator and the table were coinciding with and 0.13 mm away from the gantry isocenter, respectively. Experiments with test plans demonstrated the limited dosimetric consequences on the circular arc delivery given the aforementioned mechanical uncertainties. End-to-end measurements determined that the uncertainty of the CBCT guidance was≤1 mm in each direction with respect to the reference CT image. CONCLUSIONS:In arc delivery, the mechanical uncertainties associated with the gantry and the table do not require remarkable increases in geometric margins. If large enough, the residual setup errors following CBCT guidance will dominate the overall dosimetric consequence. Therefore, the Versa HD linac is a valid system for stereotactic radiosurgery using non-coplanar arc delivery.
Keywords: Linear accelerator, stereotactic radiosurgery, mechanical uncertainty, cone-beam CT guidance
DOI: 10.3233/XST-190580
Journal: Journal of X-Ray Science and Technology, vol. 28, no. 1, pp. 71-82, 2020
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]