Article type: Research Article
Authors: Deng, Luzhena; b | Yasar, Selcukb | Ahmed, Md Foiezb | Jayarathna, Sandunb | Feng, Penga | Wei, Biaoa | Vedantham, Srinivasanc | Karellas, Andrewc | Cho, Sang Hyunb; d; *
Affiliations: [a] Key Laboratory of Optoelectronics Technology & System, Chongqing University, Ministry of Education, Chongqing, China | [b] Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA | [c] Department of Medical Imaging, University of Arizona, Tucson, AZ, USA | [d] Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA
Correspondence: [*] Corresponding author: Sang Hyun Cho. E-mail: [email protected].
Abstract: OBJECTIVE:To investigate the image quality and x-ray dose associated with a transmission computed tomography (CT) component implemented within the same platform of an experimental benchtop x-ray fluorescence CT (XFCT) system for multimodal preclinical imaging applications. METHODS:Cone-beam CT scans were performed using an experimental benchtop CT + XFCT system and a cylindrically-shaped 3D-printed polymethyl methacrylate phantom (3 cm in diameter, 7 cm in height) loaded with various concentrations (0.05–1 wt. %) of gold nanoparticles (GNPs). Two commercial CT quality assurance phantoms containing 3D line-pair (LP) targets and contrast targets were also scanned. The x-ray beams of 40 and 62 kVp, both filtered by 0.08 mm Cu and 0.4 mm Al, were used with 17 ms of exposure time per projection at three current settings (2.5, 5, and 10 mA). The ordered-subset simultaneous algebraic reconstruction and total variation–minimization methods were used to reconstruct images. Sparse projection and short scan were considered to reduce the x-ray dose. The contrast-to-noise ratio (CNR) and modulation transfer function (MTF) were calculated. RESULTS:The lowest detectable concentration of GNPs (CNR > 5) and the highest spatial resolution (per MTF50%) were 0.10 wt. % and 9.5 LP/CM, respectively, based on the images reconstructed from 360 projections of the 40 kVp beam (or x-ray dose of 3.44 cGy). The background noise for the image resulting in the lowest GNP detection limit was 25 Hounsfield units. CONCLUSION:The transmission CT component within the current experimental benchtop CT + XFCT system produced images deemed acceptable for multimodal (CT + XFCT) imaging purposes, with less than 4 cGy of x-ray dose.
Keywords: Computed tomography, x-ray fluorescence computed tomography, x-ray dose, image resolution
DOI: 10.3233/XST-180457
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 3, pp. 431-442, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]