Abstract: In this study, we designed mobile X-ray equipment that generates high-power X-rays, using an internal power source by means of a super-capacitor, and evaluated its safety. The proposed X-ray equipment uses the charging voltage of a battery to store high density energy, supplementing the electric charge of the super-capacitor, which can instantly release a large amount of energy. Further, pulse frequency modulation was applied to produce high voltage and thereby improve energy efficiency. The developed mobile X-ray equipment enables to generate an output of 30 kW and, therefore, can be applied to many diagnostic fields. In addition, various devices and control circuits were employed to ensure convenience and safety of using the equipment in clinical applications. This study analyzed the error ranges regarding tube voltage, tube current, irradiation time, coefficient variation, half-value layer, and the output characteristics. The results showed that the proposed X-ray equipment was able to generate 800mR X-ray power under the condition of 30 kW. The coefficient variation was less than 0.05 at all measurement points, which indicates that it is possible to generate the equal amount of X-ray when the driving conditions are same. Results also showed 51.25% of transmittance at 3.5mmAL in the case of the wire, which is thicker than a common reference of 2.3mmAL and indicates that this new mobile equipment is possible to generate X-rays with relatively high permeability. In conclusion, the findings in this study suggest that the new equipment can generate consistent high-power X-rays and, therefore, can be used safely by minimizing unnecessary re-taking of images and radiation exposure.
Keywords: Mobile X-ray, super-capacitor, coefficient variation, half-value layer
DOI: 10.3233/XST-180432
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 1, pp. 131-138, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]