Article type: Research Article
Authors: Price, Samantha G.; | Silvius, Alexander A. | Izaguirre, Enrique W.;
Affiliations: Radiation Oncology, Washington University, St. Louis, MO, USA | Nuclear Science and Engineering Institute, University of MO, Columbia, MO, USA | Radiation Oncology, Baylor Scott and White Health, Temple, TX, USA
Note: [] Corresponding author: Enrique W. Izaguirre, 2401 South 31st Street, MS-09-C011, Temple, TX 76508, USA. Tel.: +1 254 724 0441; Fax: +1 254 724 6061; E-mail: [email protected]
Abstract: Image guided conformal small animal orthovoltage microirradiators are currently under development to perform radiobiological experiments with preclinical cancer models. An important component of these instruments is the treatment delivery image guidance system, a microelectronic portal imaging device (μEPID). Here, we present the design and implementation of a μEPID, specifically designed and constructed for small animal orthovoltage microirradiators. The μEPID can acquire images in the range of 60 kVp to 320 kVp x-ray photon energies and can endure high doses from orthovoltage beams without radiation damage. The μEPID can acquire 200 μm resolution images at a rate of 17 frames per second for online in vivo co-registration between irradiation beams and small animal anatomy. An exposure with less than 1% of a 2 Gy treatment field is required for imaging, which is an adequate ratio between imaging dose and treatment dose to avoid undesired irradiation of healthy tissue or alteration of the preclinical cancer model. The μEPID was calibrated for microdosimetry with a precision of 4.1% with respect to an ion chamber, used as a gold standard. To validate the in vivo device performance, irradiations of lung, brain, and xenograft breast cancer preclinical models were performed and analyzed.
Keywords: Preclinical radiobiology, microirradiator, high resolution imaging, orthovoltage imaging
DOI: 10.3233/XST-140445
Journal: Journal of X-Ray Science and Technology, vol. 22, no. 5, pp. 551-567, 2014
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]