Article type: Research Article
Authors: Zhou, Yifang; ; | Scott II, Alexander | Allahverdian, Janet | Frankel, Steve
Affiliations: Department of Radiation Safety, Cedars-Sinai Medical Center, Los Angeles, CA, USA | Department of Imaging, Cedars-Sinai Medical Center, Los Angeles, CA, USA | Department of Biomedical Sciences, Cedars-Sinai Medical Center, Los Angeles, CA, USA
Note: [] Corresponding author: Yifang Zhou, Radiation Safety and Imaging, TSB 190, Cedars-Sinai Medical Center, Los Angeles, CA 90048, USA. E-mail: [email protected]
Abstract: To quantify the trade-offs of dose and image quality among pre-loaded automatic exposure control (AEC) options in digital mammography, two AEC tables from the Hologic Selenia digital mammography system were compared: the default AEC "table 0" and AEC "table 1". Realistically-shaped phantoms consisting of tissue-equivalent material of various thicknesses (4.5 cm–7 cm) were imaged to obtain a figure of merit (FOM), the squared contrast-to-noise ratio per mean glandular dose. To relate the results to pathological findings and to evaluate the overall performance, the measured contrast-to-noise ratios were applied to simulated lesions on the anthropomorphic breast phantom images, producing various lesion configurations which were blindly scored. It was found that the AEC table 1 improves the low contrast FOM by 11% to 20% for the breast thicknesses of 4.5–6 cm. However, for the 7 cm thick breast, the AEC table 1 decreases the low contrast FOM by 17%. For microcalcifications, the AEC table 1 improves the FOM slightly for the breast thicknesses of 4.5--6 cm and decreases it by 18% at a thickness of 7 cm. The lesion simulation showed enhanced contrast due to the AEC table 1 for the breast thicknesses of 5 cm, 6 cm, and 7 cm, but the enhancement gradually reduces as the thickness increases. The lesion reading showed that the microcalcification detection was scored significantly higher from the AEC table 1 for the thicknesses 5 cm, 6 cm, and 7 cm. The corresponding improvement of mass detection scores was also observed but not consistently significant over the thickness range.
Keywords: Automatic exposure control, digital mammography, figure of merit, mean glandular dose, lesion simulation
DOI: 10.3233/XST-140433
Journal: Journal of X-Ray Science and Technology, vol. 22, no. 3, pp. 377-394, 2014
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]