Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Ji, Changguo
Affiliations: The City Key Lab of Medical Physics and Engineering, Peking University, Beijing, P.R. China. E-mail: [email protected]
Abstract: Purpose: In circular cone-beam microcomputed tomography (micro-CT), it is likely that the length of the Field of View (FOV) of a single acquisition is shorter than the total length of the object to be imaged, such as a rat in the case of preclinical application of micro-CT. This leads to multiple acquisitions using different bed positions with bed translations in between, which can be automated using a motorized bed stage. However, subtle mechanical inaccuracies can cause undesired effects when the reconstructed volumes of the different acquisitions are combined into one larger volume. In this paper, we develop an automated method for accurately stitching 3D computed tomography (CT) data using an image registration scheme, and validate this technique in a circular cone-beam micro-CT scanner. Methods: The approach is based on precalculated spatial transformation matrices acquired by a calibration phantom with point markers at stitching positions. The spatial transformation between two adjacent subvolumes was calculated only once with a rigid-body matching algorithm. Once all transformation matrices are obtained, all subsequent reconstructed subvolumes imaged at these fixed positions can be stitched accurately, and efficiently using these precalculated matrices. Results: We applied this method to real object/animal imaging in circular cone-beam micro-CT and compared the result with that obtained by stitching method calculated only by translation distances and CT voxel size. Both stitching errors calculated using point markers and stitched volumes of rigid object (a syringe) and small animal (a rat) illustrated the success of our proposed approach. Conclusions: Preliminary experimental results demonstrate that "3D data stitching" using an image registration scheme provides a good solution to the voxel mismatch caused by limited FOV length in circular cone-beam micro-CT. This method can be extended to other tomography techniques which need to acquire data at fixed scanning positions.
DOI: 10.3233/XST-2010-0246
Journal: Journal of X-Ray Science and Technology, vol. 18, no. 2, pp. 99-110, 2010
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]