Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Benítez, Ricardo Betancourt; | Ning, Ruola | Conover, David | Liu, Shaohua
Affiliations: Department of Imaging Sciences, University of Rochester Medical Center, Rochester, NY, USA | Department of Physics and Astronomy, University of Rochester, Rochester, NY, USA
Note: [] Corresponding author. E-mail: [email protected]
Abstract: The physical performance of two Flat Panel Detectors has been evaluated. The first Flat Panel Detector is for Fluoroscopic applications, Varian PaxScan 2520, and the second is for Cone Beam Computer Tomography applications, Varian PaxScan 4030CB. First, the spectrum of the X-ray source was measured. Second, the linearity of the detectors was investigated by using an ionization chamber and the average ADU values of the detectors. Third, the temporal resolution was characterized by evaluating their image lag. Fourth, their spatial resolution was characterized by the pre-sampling Modulation Transfer Function. Fifth, the Normalized Noise Power Spectrum was calculated for various exposures levels. Finally, the Detective Quantum Efficiency was obtained as a function of spatial frequency and entrance exposure. The results illustrate that the physical performance in Detective Quantum Efficiency and Normalized Noise Power Spectrum of the Cone Beam Computer Tomography detector is superior to that of the fluoroscopic detector whereas the latter detector has a higher spatial resolution as demonstrated by larger values of its Modulation Transfer Function at large spatial frequencies.
Keywords: Cone Beam Computer Tomography, Modulation Transfer Function, Edge Spread Function, Normalized Noise Power Spectrum, Detective Quantum Efficiency, Fluoroscopy Flat Panel Detector, Cone Beam Flat Panel Detector
DOI: 10.3233/XST-2009-0234
Journal: Journal of X-Ray Science and Technology, vol. 17, no. 4, pp. 279-293, 2009
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]