Article type: Research Article
Authors: Ali, Imad | Ahmad, Salahuddin | Joel, Suresh | Williamson, Jeffrey F.
Affiliations: Department of Radiation Oncology, Oklahoma University Health Sciences Center, Oklahoma City, OK 73104, USA | Department of Biomedical Engineering, Virginia Commonwealth University, Richmond, Virginia 23284-3067, USA | Department of Radiation Oncology, Virginia Commonwealth University, Richmond, VA 23298, USA
Note: [] Corresponding author. Imad Ali, Ph.D., Assistant Professor, Medical Physics, Radiation Oncology Department, Oklahoma University Health Sciences Center, 825 N.E. 10th Street, OUPB 1430, Oklahoma City, OK 73104, USA. Tel: +1 405 271 8290; Fax: +1 405 271 9240; E-mail: [email protected]
Abstract: It is well known that optical density (OD) of the radiochromic film (RCF) continues to grow after exposure at rates that have a complex dependence on dose, temperature, and densitometry wavelength. Dose rate and fractionation artifacts associated with variations in OD growth may limit the accuracy achievable by RCF dosimetry in brachytherapy and external beam applications, particularly at low doses (<5 Gy) and low dose rates (<10 cGy/h) where OD growth and sensitivity effects are large. To identify densitometry wavelengths that minimize OD growth artifacts and enhance RCF sensitivity at low doses, we have investigated Model MD-55-2 RCF response as a function of densitometry wavelength, irradiation-to-densitometry time interval, dose and temperature. Using a Perkin Elmer spectrophotometer, the absorption spectrum in the 500–700 nm range was measured for doses ranging from 1–100 Gy, over post-irradiation times from 1 h to 60 days. An empirical model with time-independent, fast and slow growth components was used to fit single exposure data and the dependence of the resulting best-fit parameters on dose and densitometry wavelength was investigated. RCF OD variation with temperature in the range 22–40� was measured. Wavelengths in the 660–690 nm range were found to minimize the dose-dependence of OD post-exposure growth. Densitometry wavelengths in the range of 670–680 nm enhance RCF sensitivity and show small variations in OD with temperature in the range from 22–40�. Compared to 633 nm light, 675 nm densitometry reduces OD growth at 1 Gy from 70% to 10% over a period of nearly 1174.0 h relative to the initial OD measured at 1.7 h post-irradiation. In addition, RCF sensitivity is nearly doubled at this wavelength for all dose levels.
DOI: 10.3233/XST-2009-0212
Journal: Journal of X-Ray Science and Technology, vol. 17, no. 1, pp. 61-73, 2009
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]