Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Zhu, Xiaohua | Yu, Jianqun | Huang, Zheng
Affiliations: Shanghai Pneumology Hospital, Shanghai, P.R. China | The 1st Affiliation Hospital of Hua-Xi Medical Center, Shichuan, P.R. China | HealthONE Alliance, Denver, Colorado, USA
Note: [] Correspondence to: Zheng Huang, HealthONE Alliance, 5125 Stapleton Drive North, Denver, CO 80216, USA. Tel.: +1 303 3204061; Fax: +1 303 3204061; E-mail: [email protected]
Abstract: The aim of this study was to optimize CT scan protocol to minimize patient exposure to ionizing radiation while maintaining CT image quality for detecting pulmonary diseases. The absorbed dose and CT scan quality were determined using dose phantoms at various tube current settings (7.5–115 mAs). The relationship between the absorbed radiation dose and CT scan quality were examined. For both 8-mm and 3-mm scanned slice thickness, the homogeneity of CT scans was within the required range (< 4 HU) at tested tube settings (> 7.5 mAs). However, the noise level of CT scans remained within the acceptance range (< 0.35%) only when the tube current was ≥ 25 mAs. The reduction of tube current from 115 mAs to 40, 25 or 7.5 mAs could decrease the CT Dose Index (CTDI) to 60, 70 or 85%, respectively. The evaluation of CT image quality suggested that there was an inflection point at 25 mAs. When the spiral chest CT scan was employed at a low tube current setting (i.e. 40 or 25 mAs), it maintained satisfactory image quality and reduced the absorbed dose of the patient, therefore protecting patients from ionizing radiation exposure.
Keywords: spiral CT scan, chest CT, low-dose, absorbed dose, radiographic exposure, radiation protection
Journal: Journal of X-Ray Science and Technology, vol. 12, no. 1, pp. 27-33, 2004
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
[email protected]
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office [email protected]
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
[email protected]
For editorial issues, like the status of your submitted paper or proposals, write to [email protected]
如果您在出版方面需要帮助或有任何建, 件至: [email protected]